Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet

INTRODUCTION: The helmet may be an effective interface for the delivery of noninvasive positive pressure ventilation. The high internal gas volume of the helmet can act as a 'mixing chamber', in which the humidity of the patient's expired alveolar gases increases the humidity of the d...

Descripción completa

Detalles Bibliográficos
Autores principales: Chiumello, Davide, Chierichetti, Monica, Tallarini, Federica, Cozzi, Paola, Cressoni, Massimo, Polli, Federico, Colombo, Riccardo, Castelli, Antonio, Gattinoni, Luciano
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2447610/
https://www.ncbi.nlm.nih.gov/pubmed/18426561
http://dx.doi.org/10.1186/cc6875
_version_ 1782156979187220480
author Chiumello, Davide
Chierichetti, Monica
Tallarini, Federica
Cozzi, Paola
Cressoni, Massimo
Polli, Federico
Colombo, Riccardo
Castelli, Antonio
Gattinoni, Luciano
author_facet Chiumello, Davide
Chierichetti, Monica
Tallarini, Federica
Cozzi, Paola
Cressoni, Massimo
Polli, Federico
Colombo, Riccardo
Castelli, Antonio
Gattinoni, Luciano
author_sort Chiumello, Davide
collection PubMed
description INTRODUCTION: The helmet may be an effective interface for the delivery of noninvasive positive pressure ventilation. The high internal gas volume of the helmet can act as a 'mixing chamber', in which the humidity of the patient's expired alveolar gases increases the humidity of the dry medical gases, thus avoiding the need for active humidification. We evaluated the temperature and humidity of respiratory gases inside the helmet, with and without a heated humidifier, during continuous positive airway pressure (CPAP) delivered with a helmet. METHODS: Nine patients with acute respiratory failure (arterial oxygen tension/fractional inspired oxygen ratio 209 ± 52 mmHg) and 10 healthy individuals were subjected to CPAP. The CPAP was delivered either through a mechanical ventilator or by continuous low (40 l/min) or high flow (80 l/min). Humidity was measured inside the helmet using a capacitive hygrometer. The level of patient comfort was evaluated using a continuous scale. RESULTS: In patients with acute respiratory failure, the heated humidifier significantly increased the absolute humidity from 18.4 ± 5.5 mgH(2)O/l to 34.1 ± 2.8 mgH(2)O/l during ventilator CPAP, from 11.4 ± 4.8 mgH(2)O/l to 33.9 ± 1.9 mgH(2)O/l during continuous low-flow CPAP, and from 6.4 ± 1.8 mgH(2)O/l to 24.2 ± 5.4 mgH(2)O/l during continuous high-flow CPAP. Without the heated humidifier, the absolute humidity was significantly higher with ventilator CPAP than with continuous low-flow and high-flow CPAP. The level of comfort was similar for all the three modes of ventilation and with or without the heated humidifier. The findings in healthy individuals were similar to those in the patients with acute respiratory failure. CONCLUSION: The fresh gas flowing through the helmet with continuous flow CPAP systems limited the possibility to increase the humidity. We suggest that a heated humidifier should be employed with continuous flow CPAP systems.
format Text
id pubmed-2447610
institution National Center for Biotechnology Information
language English
publishDate 2008
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-24476102008-07-10 Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet Chiumello, Davide Chierichetti, Monica Tallarini, Federica Cozzi, Paola Cressoni, Massimo Polli, Federico Colombo, Riccardo Castelli, Antonio Gattinoni, Luciano Crit Care Research INTRODUCTION: The helmet may be an effective interface for the delivery of noninvasive positive pressure ventilation. The high internal gas volume of the helmet can act as a 'mixing chamber', in which the humidity of the patient's expired alveolar gases increases the humidity of the dry medical gases, thus avoiding the need for active humidification. We evaluated the temperature and humidity of respiratory gases inside the helmet, with and without a heated humidifier, during continuous positive airway pressure (CPAP) delivered with a helmet. METHODS: Nine patients with acute respiratory failure (arterial oxygen tension/fractional inspired oxygen ratio 209 ± 52 mmHg) and 10 healthy individuals were subjected to CPAP. The CPAP was delivered either through a mechanical ventilator or by continuous low (40 l/min) or high flow (80 l/min). Humidity was measured inside the helmet using a capacitive hygrometer. The level of patient comfort was evaluated using a continuous scale. RESULTS: In patients with acute respiratory failure, the heated humidifier significantly increased the absolute humidity from 18.4 ± 5.5 mgH(2)O/l to 34.1 ± 2.8 mgH(2)O/l during ventilator CPAP, from 11.4 ± 4.8 mgH(2)O/l to 33.9 ± 1.9 mgH(2)O/l during continuous low-flow CPAP, and from 6.4 ± 1.8 mgH(2)O/l to 24.2 ± 5.4 mgH(2)O/l during continuous high-flow CPAP. Without the heated humidifier, the absolute humidity was significantly higher with ventilator CPAP than with continuous low-flow and high-flow CPAP. The level of comfort was similar for all the three modes of ventilation and with or without the heated humidifier. The findings in healthy individuals were similar to those in the patients with acute respiratory failure. CONCLUSION: The fresh gas flowing through the helmet with continuous flow CPAP systems limited the possibility to increase the humidity. We suggest that a heated humidifier should be employed with continuous flow CPAP systems. BioMed Central 2008 2008-04-21 /pmc/articles/PMC2447610/ /pubmed/18426561 http://dx.doi.org/10.1186/cc6875 Text en Copyright © 2008 Chiumello et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an open access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Chiumello, Davide
Chierichetti, Monica
Tallarini, Federica
Cozzi, Paola
Cressoni, Massimo
Polli, Federico
Colombo, Riccardo
Castelli, Antonio
Gattinoni, Luciano
Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet
title Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet
title_full Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet
title_fullStr Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet
title_full_unstemmed Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet
title_short Effect of a heated humidifier during continuous positive airway pressure delivered by a helmet
title_sort effect of a heated humidifier during continuous positive airway pressure delivered by a helmet
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2447610/
https://www.ncbi.nlm.nih.gov/pubmed/18426561
http://dx.doi.org/10.1186/cc6875
work_keys_str_mv AT chiumellodavide effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT chierichettimonica effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT tallarinifederica effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT cozzipaola effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT cressonimassimo effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT pollifederico effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT colomboriccardo effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT castelliantonio effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet
AT gattinoniluciano effectofaheatedhumidifierduringcontinuouspositiveairwaypressuredeliveredbyahelmet