A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation
BACKGROUND: Non-operative management of midshaft clavicle fractures is standard; however, surgical management is increasing. The purpose of this study is to compare the biomechanical performance of plate versus intramedullary fixation in cyclic bending for matched pairs of cadaveric clavicles. We hy...
Autores principales: | , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2488335/ https://www.ncbi.nlm.nih.gov/pubmed/18631380 http://dx.doi.org/10.1186/1749-799X-3-28 |
_version_ | 1782158118091751424 |
---|---|
author | Golish, S Raymond Oliviero, Jason A Francke, Eric I Miller, Mark D |
author_facet | Golish, S Raymond Oliviero, Jason A Francke, Eric I Miller, Mark D |
author_sort | Golish, S Raymond |
collection | PubMed |
description | BACKGROUND: Non-operative management of midshaft clavicle fractures is standard; however, surgical management is increasing. The purpose of this study is to compare the biomechanical performance of plate versus intramedullary fixation in cyclic bending for matched pairs of cadaveric clavicles. We hypothesized that the biomechanical properties are similar. METHODS: Eight sets of matched clavicles with vertical, midshaft osteotomies were prepared from fresh, frozen cadavers. A 3.5 mm dynamic compression plate or a 3.8 or 4.5 mm intramedullary device were used for fixation. Clavicles were loaded in a four-point bend at 6 different loads for 3000 cycles at 1 Hz starting with 180 N and increasing by 180 N with sampling at 2 Hz. Failure was defined as 10 mm of displacement or catastrophic construct failure prior to 10 mm of displacement. RESULTS: Between constructs, there was a significant difference with large effect size in displacement at fixed loads of 180 N (P = 0.001; Cohen's d = 1.85), 360 N (P = 0.033; Cohen's d = 1.39), 540 N (P = 0.003; Cohen's d = 0.73) and 720 N (P = 0.018; Cohen's d = 0.72). There was a significant difference with large effect size in load at fixed displacements of 5 mm (P = 0.001; Cohen's d = 1.49), 7.5 mm (P = 0.011; Cohen's d = 1.06), and 10 mm (P = 0.026; Cohen's d = 0.84). CONCLUSION: Plate constructs are superior in showing less displacement at fixed loads and greater loads at fixed displacements over a broad range of loads and displacements with cyclic four-point bending. The clinical relevance is that plate fixation may provide a stronger construct for early rehabilitation protocols that focus on repetitive movements in the early pre-operative period. |
format | Text |
id | pubmed-2488335 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-24883352008-07-29 A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation Golish, S Raymond Oliviero, Jason A Francke, Eric I Miller, Mark D J Orthop Surg Research Article BACKGROUND: Non-operative management of midshaft clavicle fractures is standard; however, surgical management is increasing. The purpose of this study is to compare the biomechanical performance of plate versus intramedullary fixation in cyclic bending for matched pairs of cadaveric clavicles. We hypothesized that the biomechanical properties are similar. METHODS: Eight sets of matched clavicles with vertical, midshaft osteotomies were prepared from fresh, frozen cadavers. A 3.5 mm dynamic compression plate or a 3.8 or 4.5 mm intramedullary device were used for fixation. Clavicles were loaded in a four-point bend at 6 different loads for 3000 cycles at 1 Hz starting with 180 N and increasing by 180 N with sampling at 2 Hz. Failure was defined as 10 mm of displacement or catastrophic construct failure prior to 10 mm of displacement. RESULTS: Between constructs, there was a significant difference with large effect size in displacement at fixed loads of 180 N (P = 0.001; Cohen's d = 1.85), 360 N (P = 0.033; Cohen's d = 1.39), 540 N (P = 0.003; Cohen's d = 0.73) and 720 N (P = 0.018; Cohen's d = 0.72). There was a significant difference with large effect size in load at fixed displacements of 5 mm (P = 0.001; Cohen's d = 1.49), 7.5 mm (P = 0.011; Cohen's d = 1.06), and 10 mm (P = 0.026; Cohen's d = 0.84). CONCLUSION: Plate constructs are superior in showing less displacement at fixed loads and greater loads at fixed displacements over a broad range of loads and displacements with cyclic four-point bending. The clinical relevance is that plate fixation may provide a stronger construct for early rehabilitation protocols that focus on repetitive movements in the early pre-operative period. BioMed Central 2008-07-16 /pmc/articles/PMC2488335/ /pubmed/18631380 http://dx.doi.org/10.1186/1749-799X-3-28 Text en Copyright © 2008 Golish et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Golish, S Raymond Oliviero, Jason A Francke, Eric I Miller, Mark D A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation |
title | A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation |
title_full | A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation |
title_fullStr | A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation |
title_full_unstemmed | A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation |
title_short | A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation |
title_sort | biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2488335/ https://www.ncbi.nlm.nih.gov/pubmed/18631380 http://dx.doi.org/10.1186/1749-799X-3-28 |
work_keys_str_mv | AT golishsraymond abiomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation AT olivierojasona abiomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation AT franckeerici abiomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation AT millermarkd abiomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation AT golishsraymond biomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation AT olivierojasona biomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation AT franckeerici biomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation AT millermarkd biomechanicalstudyofplateversusintramedullarydevicesformidshaftclaviclefixation |