Cargando…

Tono-Pen XL tonometry during application of a suction ring in rabbits

BACKGROUND: The purpose of this study is to evaluate the use of Tono-Pen XL in measuring IOP during the application of a suction ring in rabbit eyes with manometrically controlled IOP. METHODS: Tono-Pen XL was calibrated against direct manometry in 10 rabbit eyes. A suction ring was then applied in...

Descripción completa

Detalles Bibliográficos
Autores principales: Charisis, Spyridon K, Ginis, Harilaos S, Kounis, Georgios A, Tsilimbaris, Miltiadis K
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2515826/
https://www.ncbi.nlm.nih.gov/pubmed/18667058
http://dx.doi.org/10.1186/1471-2415-8-14
Descripción
Sumario:BACKGROUND: The purpose of this study is to evaluate the use of Tono-Pen XL in measuring IOP during the application of a suction ring in rabbit eyes with manometrically controlled IOP. METHODS: Tono-Pen XL was calibrated against direct manometry in 10 rabbit eyes. A suction ring was then applied in 4 rabbit eyes and the IOP was determined manometrically during suction ring application at 350 mmHg vacuum pressure. Finally, in 6 catheterized rabbit eyes the IOP was measured with Tono-Pen XL during suction ring application at suction vacuum from 350 to 650 mmHg, while keeping actual IOP stable at 30 mmHg and 60 mmHg. RESULTS: Linear regression analysis revealed that the Tono-pen XL was reliable for IOPs between 10 and 70 mmHg (R(2 )= 0.9855). Direct manometry during suction ring application showed no statistically significant variation of Tono-Pen XL readings when the incanulation manometry intraocular pressure changed from 30 mmHg to 60 mmHg and no statistically significant correlation between suction vacuum and IOP measurements. CONCLUSION: Tono-Pen XL measurements are unreliable during the application of a suction ring on living rabbit eyes even when the actual IOP is forced to be within the validated range of Tono-Pen XL measurements. This inaccuracy is probably related to altered corneal and scleral geometry and stress.