The Kemtek 1000 Sample Processor has been evaluated for precision, accuracy, speed and reliability. Precision was better than 1.0% at all volumes tested and accuracy within ±5%. A l00-tube assay could be set up within 15 min when patient specimens plus two reagents were sampled using a two probe sys...

Descripción completa

Detalles Bibliográficos
Autores principales: Wheeler, M. J., Waiters, Linzi
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 1988
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547724/
https://www.ncbi.nlm.nih.gov/pubmed/18925187
http://dx.doi.org/10.1155/S1463924688000100
_version_ 1782159324985950208
author Wheeler, M. J.
Waiters, Linzi
author_facet Wheeler, M. J.
Waiters, Linzi
author_sort Wheeler, M. J.
collection PubMed
description The Kemtek 1000 Sample Processor has been evaluated for precision, accuracy, speed and reliability. Precision was better than 1.0% at all volumes tested and accuracy within ±5%. A l00-tube assay could be set up within 15 min when patient specimens plus two reagents were sampled using a two probe system. Carry-over could be reduced to <0.01% by using a sufficient number of wash steps, the latter being related to the assay requirements. Evidence was found for adsorption of protein to the probe tubing but inaccuracies due to this could be reduced by introducing wash steps between samples. Problems over 12 months have been minor and quickly resolved. The authors were pleased with the way the processor performed and their staffhave confidence in leaving it to set up their assays.
format Text
id pubmed-2547724
institution National Center for Biotechnology Information
language English
publishDate 1988
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-25477242008-10-16 An evaluation of the Kemtek 1000 sample processor Wheeler, M. J. Waiters, Linzi J Automat Chem Research Article The Kemtek 1000 Sample Processor has been evaluated for precision, accuracy, speed and reliability. Precision was better than 1.0% at all volumes tested and accuracy within ±5%. A l00-tube assay could be set up within 15 min when patient specimens plus two reagents were sampled using a two probe system. Carry-over could be reduced to <0.01% by using a sufficient number of wash steps, the latter being related to the assay requirements. Evidence was found for adsorption of protein to the probe tubing but inaccuracies due to this could be reduced by introducing wash steps between samples. Problems over 12 months have been minor and quickly resolved. The authors were pleased with the way the processor performed and their staffhave confidence in leaving it to set up their assays. Hindawi Publishing Corporation 1988 /pmc/articles/PMC2547724/ /pubmed/18925187 http://dx.doi.org/10.1155/S1463924688000100 Text en Copyright © 1988 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Wheeler, M. J.
Waiters, Linzi
An evaluation of the Kemtek 1000 sample processor
title An evaluation of the Kemtek 1000 sample processor
title_full An evaluation of the Kemtek 1000 sample processor
title_fullStr An evaluation of the Kemtek 1000 sample processor
title_full_unstemmed An evaluation of the Kemtek 1000 sample processor
title_short An evaluation of the Kemtek 1000 sample processor
title_sort evaluation of the kemtek 1000 sample processor
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547724/
https://www.ncbi.nlm.nih.gov/pubmed/18925187
http://dx.doi.org/10.1155/S1463924688000100
work_keys_str_mv AT wheelermj anevaluationofthekemtek1000sampleprocessor
AT waiterslinzi anevaluationofthekemtek1000sampleprocessor
AT wheelermj evaluationofthekemtek1000sampleprocessor
AT waiterslinzi evaluationofthekemtek1000sampleprocessor