Cargando…

A novel approach to non-segmented flow analysis. Part 2. A prototype high-performance analyser

A high-performance continuous flow analyser is described, based on gas pressure driven carrier and reagents controlled by computer switched solenoid valves. The principal characteristics of the analyser are discussed and examples of its performance are provided in the form of results obtained using...

Descripción completa

Detalles Bibliográficos
Autores principales: Malcolme-Lawes, D. J., Pasquini, C.
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 1988
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547734/
https://www.ncbi.nlm.nih.gov/pubmed/18925185
http://dx.doi.org/10.1155/S1463924688000082
Descripción
Sumario:A high-performance continuous flow analyser is described, based on gas pressure driven carrier and reagents controlled by computer switched solenoid valves. The principal characteristics of the analyser are discussed and examples of its performance are provided in the form of results obtained using a standard procedure for the determination of Cr(VI). The system was also tested in use with real samples using an ammonium ion analysis on potable and effluent water samples, and the results compared with those obtained using a segmented continuous flow method operated at the Laboratory of the Government Chemist.