Cargando…
An evaluation of osmolality measurement by freezing point depression using micro-amounts of sample
An evaluation of the Advanced micro-osmometer is presented. This instrument has been shown to have an excellent analytical precision (within-run CV = 0.59%, between-day CV = 0.58%). It is accurate over an analytical range of 0-2000 mmol/kg of osmolality shown by linearity studies and split sample co...
Autores principales: | Koumantakis, G., Wyndham, L. E. |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
1989
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547790/ https://www.ncbi.nlm.nih.gov/pubmed/18925239 http://dx.doi.org/10.1155/S1463924689000167 |
Ejemplares similares
-
Importance of sample volume to the measurement and interpretation of plasma osmolality
por: Sollanek, Kurt J., et al.
Publicado: (2018) -
Evaluation of conductivity‐based osmolality measurement in urine using the Sysmex UF5000
por: Yoo, Dong‐Won, et al.
Publicado: (2020) -
Freezing Point Depression of Plasma
por: Meschia, Giacomo, et al.
Publicado: (1956) -
Predicting changes in osmolality
por: Yang, Zhe, et al.
Publicado: (2021) -
Airway Surface Liquid Osmolality Measured Using Fluorophore-Encapsulated Liposomes
por: Jayaraman, Sujatha, et al.
Publicado: (2001)