Cargando…
An evaluation of the Diamat HPLC analyser for simultaneous determination of haemoglobins A(2) and F
The authors describe a modification of the instrumental parameters of the Diamat fully automated HPLC system for Hb A(2) assay (Bio-Rad Laboratories, Milan, Italy) in order to obtain simultaneous determination of Hb A(2) and Hb F. Hb A(2) and Hb F measurements are reproducible (within-run CV 2.6%, w...
Autores principales: | Mosca, Andrea, Carpinelli, Assunta, Majavacca, Rita, Cantu'-Rajnoldi, Angelo, Garatti, Massimo, Paleari, Renata, Ferrari, Maurizio, Agape, Vittorio, Maccioni, Liliana, Pisano, Sandra, Galanello, Renzo |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
1989
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547811/ https://www.ncbi.nlm.nih.gov/pubmed/18925255 http://dx.doi.org/10.1155/S1463924689000520 |
Ejemplares similares
-
Evaluation of an automatic HPLC analyser for thalassemia and haemoglobin variants screening
por: Galanello, R., et al.
Publicado: (1995) -
Evaluation of a new semi-automated high-performance liquid chromatography
method for glycosylated haemoglobins
por: Carpinelli, Assunta, et al.
Publicado: (1986) -
Viability of ex-vivo myography as a diagnostic tool for rectus abdominis muscle electrical activity collected at Cesarean section within a diamater cohort study
por: Reyes, David R. A., et al.
Publicado: (2022) -
Deferiprone in the treatment of transfusion-dependent thalassemia: a review and perspective
por: Galanello, Renzo
Publicado: (2007) -
Structural and Functional Characterization of a New Double Variant Haemoglobin (HbG-Philadelphia/Duarte α
(2)
(68Asn→Lys)
β
(2)
(62Ala→Pro))
por: Fais, Antonella, et al.
Publicado: (2011)