Analytical performance of the selective, automatic multianalyser Olympus AU 5031

The analytical performance of the selective, automatic multianalyser Olympus AU 5031 was evaluated over four months and assessed for practicability for another eight months. The evaluation followed the ECCLS guidelines. Twenty routine parameters were measured. In addition, sodium and potassium were...

Descripción completa

Detalles Bibliográficos
Autores principales: Luley, C., Struss, H. G., Prellwitz, W.
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 1990
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547815/
https://www.ncbi.nlm.nih.gov/pubmed/18925258
http://dx.doi.org/10.1155/S1463924690000025
_version_ 1782159346280431616
author Luley, C.
Struss, H. G.
Prellwitz, W.
author_facet Luley, C.
Struss, H. G.
Prellwitz, W.
author_sort Luley, C.
collection PubMed
description The analytical performance of the selective, automatic multianalyser Olympus AU 5031 was evaluated over four months and assessed for practicability for another eight months. The evaluation followed the ECCLS guidelines. Twenty routine parameters were measured. In addition, sodium and potassium were determined on an attached flame photometric unit. Both the agreement between the eights photometers per unit and the temperature behaviour in the cuvettes was satisfactory. The imprecisions were very good. The within-run imprecision was below 1.5% for the majority of the parameters. The imprecision between days was below 5%, with the exception of creatine phosphokinase (7.4%). Glutamate dehydrogenase gave an imprecision of between 4.0% and 15.9%, which, however, is more likely due to the low activities measured rather than the fault of analyser. The recovery of the assigned values in 12 control sera was between 95% and 105% for 14 tests. Three of the remaining eight tests yielded recoveries with deviations between 10% and 18% (alanine aminotransferase, aspartate aminotransferase and bilirubin). No drift effects were observed and neither a sample carry-over nor a reagent carry-over were detected. Most tests were linear over a very wide range. Only afew tests (mainly lipase and glutamate dehydrogenase) required measurement repetitions with diluted samples. The correlation with routine instruments and tests was close. However, corrections were necessary for 14 of the 22 tests. This was not due to the performance of the analyser but, rather, to the different methodologies of compared tests, or different working temperatures on the comparison instruments, or a lack of accuracy for some of the AU tests.
format Text
id pubmed-2547815
institution National Center for Biotechnology Information
language English
publishDate 1990
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-25478152008-10-16 Analytical performance of the selective, automatic multianalyser Olympus AU 5031 Luley, C. Struss, H. G. Prellwitz, W. J Automat Chem Research Article The analytical performance of the selective, automatic multianalyser Olympus AU 5031 was evaluated over four months and assessed for practicability for another eight months. The evaluation followed the ECCLS guidelines. Twenty routine parameters were measured. In addition, sodium and potassium were determined on an attached flame photometric unit. Both the agreement between the eights photometers per unit and the temperature behaviour in the cuvettes was satisfactory. The imprecisions were very good. The within-run imprecision was below 1.5% for the majority of the parameters. The imprecision between days was below 5%, with the exception of creatine phosphokinase (7.4%). Glutamate dehydrogenase gave an imprecision of between 4.0% and 15.9%, which, however, is more likely due to the low activities measured rather than the fault of analyser. The recovery of the assigned values in 12 control sera was between 95% and 105% for 14 tests. Three of the remaining eight tests yielded recoveries with deviations between 10% and 18% (alanine aminotransferase, aspartate aminotransferase and bilirubin). No drift effects were observed and neither a sample carry-over nor a reagent carry-over were detected. Most tests were linear over a very wide range. Only afew tests (mainly lipase and glutamate dehydrogenase) required measurement repetitions with diluted samples. The correlation with routine instruments and tests was close. However, corrections were necessary for 14 of the 22 tests. This was not due to the performance of the analyser but, rather, to the different methodologies of compared tests, or different working temperatures on the comparison instruments, or a lack of accuracy for some of the AU tests. Hindawi Publishing Corporation 1990 /pmc/articles/PMC2547815/ /pubmed/18925258 http://dx.doi.org/10.1155/S1463924690000025 Text en Copyright © 1990 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Luley, C.
Struss, H. G.
Prellwitz, W.
Analytical performance of the selective, automatic multianalyser Olympus AU 5031
title Analytical performance of the selective, automatic multianalyser Olympus AU 5031
title_full Analytical performance of the selective, automatic multianalyser Olympus AU 5031
title_fullStr Analytical performance of the selective, automatic multianalyser Olympus AU 5031
title_full_unstemmed Analytical performance of the selective, automatic multianalyser Olympus AU 5031
title_short Analytical performance of the selective, automatic multianalyser Olympus AU 5031
title_sort analytical performance of the selective, automatic multianalyser olympus au 5031
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547815/
https://www.ncbi.nlm.nih.gov/pubmed/18925258
http://dx.doi.org/10.1155/S1463924690000025
work_keys_str_mv AT luleyc analyticalperformanceoftheselectiveautomaticmultianalyserolympusau5031
AT strusshg analyticalperformanceoftheselectiveautomaticmultianalyserolympusau5031
AT prellwitzw analyticalperformanceoftheselectiveautomaticmultianalyserolympusau5031