Evaluation of the Olympus AU-510 analyser

The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of t...

Descripción completa

Detalles Bibliográficos
Autores principales: Farré, Carmen, Velasco, Jesús, Ramón, Francisco
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 1991
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547928/
https://www.ncbi.nlm.nih.gov/pubmed/18924907
http://dx.doi.org/10.1155/S1463924691000366
_version_ 1782159365218762752
author Farré, Carmen
Velasco, Jesús
Ramón, Francisco
author_facet Farré, Carmen
Velasco, Jesús
Ramón, Francisco
author_sort Farré, Carmen
collection PubMed
description The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of the analytical units and then an evaluation in routine work conditions. The operational characteristics of the system were also studied. The first stage included a photometric study: dependent on the absorbance, the inaccuracy varies between +0.5% to -0.6% at 405 nm and from -5.6% to 10.6% at 340 nm; the imprecision ranges between -0.22% and 0.56% at 405 nm and between 0.09% and 2.74% at 340 nm. Linearity was acceptable, apart from a very low absorbance for NADH at 340 nm; and the imprecision of the serum sample pipetter was satisfactory. Twelve serum analytes were studied under routine conditions: glucose, urea urate, cholesterol, triglycerides, total bilirubin, creatinine, phosphate, iron, aspartate aminotransferase, alanine aminotransferase and gamma-glutamyl transferase. The within-run imprecision (CV%) ranged from 0.67% for phosphate to 2.89% for iron and the between-run imprecision from 0.97% for total bilirubin to 7.06% for iron. There was no carryover in a study of the serum sample pipetter. Carry-over studies with the reagent and sample pipetters shows some cross contamination in the iron assay.
format Text
id pubmed-2547928
institution National Center for Biotechnology Information
language English
publishDate 1991
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-25479282008-10-16 Evaluation of the Olympus AU-510 analyser Farré, Carmen Velasco, Jesús Ramón, Francisco J Automat Chem Research Article The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of the analytical units and then an evaluation in routine work conditions. The operational characteristics of the system were also studied. The first stage included a photometric study: dependent on the absorbance, the inaccuracy varies between +0.5% to -0.6% at 405 nm and from -5.6% to 10.6% at 340 nm; the imprecision ranges between -0.22% and 0.56% at 405 nm and between 0.09% and 2.74% at 340 nm. Linearity was acceptable, apart from a very low absorbance for NADH at 340 nm; and the imprecision of the serum sample pipetter was satisfactory. Twelve serum analytes were studied under routine conditions: glucose, urea urate, cholesterol, triglycerides, total bilirubin, creatinine, phosphate, iron, aspartate aminotransferase, alanine aminotransferase and gamma-glutamyl transferase. The within-run imprecision (CV%) ranged from 0.67% for phosphate to 2.89% for iron and the between-run imprecision from 0.97% for total bilirubin to 7.06% for iron. There was no carryover in a study of the serum sample pipetter. Carry-over studies with the reagent and sample pipetters shows some cross contamination in the iron assay. Hindawi Publishing Corporation 1991 /pmc/articles/PMC2547928/ /pubmed/18924907 http://dx.doi.org/10.1155/S1463924691000366 Text en Copyright © 1991 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Farré, Carmen
Velasco, Jesús
Ramón, Francisco
Evaluation of the Olympus AU-510 analyser
title Evaluation of the Olympus AU-510 analyser
title_full Evaluation of the Olympus AU-510 analyser
title_fullStr Evaluation of the Olympus AU-510 analyser
title_full_unstemmed Evaluation of the Olympus AU-510 analyser
title_short Evaluation of the Olympus AU-510 analyser
title_sort evaluation of the olympus au-510 analyser
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547928/
https://www.ncbi.nlm.nih.gov/pubmed/18924907
http://dx.doi.org/10.1155/S1463924691000366
work_keys_str_mv AT farrecarmen evaluationoftheolympusau510analyser
AT velascojesus evaluationoftheolympusau510analyser
AT ramonfrancisco evaluationoftheolympusau510analyser