Evaluation of the Olympus AU-510 analyser
The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of t...
Autores principales: | , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
1991
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547928/ https://www.ncbi.nlm.nih.gov/pubmed/18924907 http://dx.doi.org/10.1155/S1463924691000366 |
_version_ | 1782159365218762752 |
---|---|
author | Farré, Carmen Velasco, Jesús Ramón, Francisco |
author_facet | Farré, Carmen Velasco, Jesús Ramón, Francisco |
author_sort | Farré, Carmen |
collection | PubMed |
description | The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of the analytical units and then an evaluation in routine work conditions. The operational characteristics of the system were also studied. The first stage included a photometric study: dependent on the absorbance, the inaccuracy varies between +0.5% to -0.6% at 405 nm and from -5.6% to 10.6% at 340 nm; the imprecision ranges between -0.22% and 0.56% at 405 nm and between 0.09% and 2.74% at 340 nm. Linearity was acceptable, apart from a very low absorbance for NADH at 340 nm; and the imprecision of the serum sample pipetter was satisfactory. Twelve serum analytes were studied under routine conditions: glucose, urea urate, cholesterol, triglycerides, total bilirubin, creatinine, phosphate, iron, aspartate aminotransferase, alanine aminotransferase and gamma-glutamyl transferase. The within-run imprecision (CV%) ranged from 0.67% for phosphate to 2.89% for iron and the between-run imprecision from 0.97% for total bilirubin to 7.06% for iron. There was no carryover in a study of the serum sample pipetter. Carry-over studies with the reagent and sample pipetters shows some cross contamination in the iron assay. |
format | Text |
id | pubmed-2547928 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 1991 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-25479282008-10-16 Evaluation of the Olympus AU-510 analyser Farré, Carmen Velasco, Jesús Ramón, Francisco J Automat Chem Research Article The selective multitest Olympus AU-510 analyser was evaluated according to the recommendations of the Comision de Instrumentacion de la Sociedad Española de Quimica Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was carried out in two stages: an examination of the analytical units and then an evaluation in routine work conditions. The operational characteristics of the system were also studied. The first stage included a photometric study: dependent on the absorbance, the inaccuracy varies between +0.5% to -0.6% at 405 nm and from -5.6% to 10.6% at 340 nm; the imprecision ranges between -0.22% and 0.56% at 405 nm and between 0.09% and 2.74% at 340 nm. Linearity was acceptable, apart from a very low absorbance for NADH at 340 nm; and the imprecision of the serum sample pipetter was satisfactory. Twelve serum analytes were studied under routine conditions: glucose, urea urate, cholesterol, triglycerides, total bilirubin, creatinine, phosphate, iron, aspartate aminotransferase, alanine aminotransferase and gamma-glutamyl transferase. The within-run imprecision (CV%) ranged from 0.67% for phosphate to 2.89% for iron and the between-run imprecision from 0.97% for total bilirubin to 7.06% for iron. There was no carryover in a study of the serum sample pipetter. Carry-over studies with the reagent and sample pipetters shows some cross contamination in the iron assay. Hindawi Publishing Corporation 1991 /pmc/articles/PMC2547928/ /pubmed/18924907 http://dx.doi.org/10.1155/S1463924691000366 Text en Copyright © 1991 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Farré, Carmen Velasco, Jesús Ramón, Francisco Evaluation of the Olympus AU-510 analyser |
title | Evaluation of the Olympus AU-510 analyser |
title_full | Evaluation of the Olympus AU-510 analyser |
title_fullStr | Evaluation of the Olympus AU-510 analyser |
title_full_unstemmed | Evaluation of the Olympus AU-510 analyser |
title_short | Evaluation of the Olympus AU-510 analyser |
title_sort | evaluation of the olympus au-510 analyser |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547928/ https://www.ncbi.nlm.nih.gov/pubmed/18924907 http://dx.doi.org/10.1155/S1463924691000366 |
work_keys_str_mv | AT farrecarmen evaluationoftheolympusau510analyser AT velascojesus evaluationoftheolympusau510analyser AT ramonfrancisco evaluationoftheolympusau510analyser |