The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study
This report describes the results of a collaborative study organized by a joint working group of the IFCC and WHO and involving nine manufacturers of TSH immunometric assay kits. The study was designed to determine whether a calibrator with a common matrix gives better between-laboratory agreement f...
Autores principales: | , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
1991
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547934/ https://www.ncbi.nlm.nih.gov/pubmed/18924906 http://dx.doi.org/10.1155/S1463924691000354 |
_version_ | 1782159366674186240 |
---|---|
author | Gaines-Das, R. E. Bristow, A. F. Brettschneider, H. |
author_facet | Gaines-Das, R. E. Bristow, A. F. Brettschneider, H. |
author_sort | Gaines-Das, R. E. |
collection | PubMed |
description | This report describes the results of a collaborative study organized by a joint working group of the IFCC and WHO and involving nine manufacturers of TSH immunometric assay kits. The study was designed to determine whether a calibrator with a common matrix gives better between-laboratory agreement for calibration of serum samples than the various kit calibrators, and to assess various materials for their suitability for use as common matrices. Kit calibrators, or calibrators consisting of the IRP for TSH made up in two common matrices: (a) serum from patients with untreated thyrotoxicosis or (b) serum taken from subjects treated with suppressive doses of triiodothyronine, gave similar results for the between-laboratory variation of estimates of TSH concentration for a range of serum samples. Dose-response curves for the two calibrators in ‘common’ matrices were similar to one another and to those for the kit calibrator. However, the occurrence of non-specific serum effects is shown by the comparison of results for these calibrators with results for calibrators made up in a third common matrix: serum treated with wheat germ lectin. Dose response curves for this calibrator were dissimilar to those for the other calibrators and between-laboratory variation for estimates in terms of this latter calibrator showed a substantial increase. Moreover, although the between-laboratory variances for estimates of the TSH concentration in terms of each of these calibrators (except those made up in serum treated with the wheat germ lectin) were similar for any one sample from five hyperthyroid patients, the variances were not consistent between samples, even for samples with similar mean TSH concentrations. These results suggest that a major factor in the between-laboratory variation, especially in the region near ‘zero dose’, is sample-related, and is caused by particular samples interacting differently with different assay systems. In general, it would appear that for the well-controlled ‘ultrasensitive’ TSH immunometric assay kits, included in this study, between-laboratory agreement of estimates of the TSH concentration in serum samples is not likely to be substantially improved by use of a common matrix for the standards. |
format | Text |
id | pubmed-2547934 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 1991 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-25479342008-10-16 The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study Gaines-Das, R. E. Bristow, A. F. Brettschneider, H. J Automat Chem Research Article This report describes the results of a collaborative study organized by a joint working group of the IFCC and WHO and involving nine manufacturers of TSH immunometric assay kits. The study was designed to determine whether a calibrator with a common matrix gives better between-laboratory agreement for calibration of serum samples than the various kit calibrators, and to assess various materials for their suitability for use as common matrices. Kit calibrators, or calibrators consisting of the IRP for TSH made up in two common matrices: (a) serum from patients with untreated thyrotoxicosis or (b) serum taken from subjects treated with suppressive doses of triiodothyronine, gave similar results for the between-laboratory variation of estimates of TSH concentration for a range of serum samples. Dose-response curves for the two calibrators in ‘common’ matrices were similar to one another and to those for the kit calibrator. However, the occurrence of non-specific serum effects is shown by the comparison of results for these calibrators with results for calibrators made up in a third common matrix: serum treated with wheat germ lectin. Dose response curves for this calibrator were dissimilar to those for the other calibrators and between-laboratory variation for estimates in terms of this latter calibrator showed a substantial increase. Moreover, although the between-laboratory variances for estimates of the TSH concentration in terms of each of these calibrators (except those made up in serum treated with the wheat germ lectin) were similar for any one sample from five hyperthyroid patients, the variances were not consistent between samples, even for samples with similar mean TSH concentrations. These results suggest that a major factor in the between-laboratory variation, especially in the region near ‘zero dose’, is sample-related, and is caused by particular samples interacting differently with different assay systems. In general, it would appear that for the well-controlled ‘ultrasensitive’ TSH immunometric assay kits, included in this study, between-laboratory agreement of estimates of the TSH concentration in serum samples is not likely to be substantially improved by use of a common matrix for the standards. Hindawi Publishing Corporation 1991 /pmc/articles/PMC2547934/ /pubmed/18924906 http://dx.doi.org/10.1155/S1463924691000354 Text en Copyright © 1991 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Gaines-Das, R. E. Bristow, A. F. Brettschneider, H. The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study |
title | The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study |
title_full | The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study |
title_fullStr | The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study |
title_full_unstemmed | The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study |
title_short | The effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for TSH: Report of a joint WHO/IFCC collaborative study |
title_sort | effects of common matrices for assay standards on performance of ‘ultra sensitive’ immunometric assays for tsh: report of a joint who/ifcc collaborative study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2547934/ https://www.ncbi.nlm.nih.gov/pubmed/18924906 http://dx.doi.org/10.1155/S1463924691000354 |
work_keys_str_mv | AT gainesdasre theeffectsofcommonmatricesforassaystandardsonperformanceofultrasensitiveimmunometricassaysfortshreportofajointwhoifcccollaborativestudy AT bristowaf theeffectsofcommonmatricesforassaystandardsonperformanceofultrasensitiveimmunometricassaysfortshreportofajointwhoifcccollaborativestudy AT brettschneiderh theeffectsofcommonmatricesforassaystandardsonperformanceofultrasensitiveimmunometricassaysfortshreportofajointwhoifcccollaborativestudy AT gainesdasre effectsofcommonmatricesforassaystandardsonperformanceofultrasensitiveimmunometricassaysfortshreportofajointwhoifcccollaborativestudy AT bristowaf effectsofcommonmatricesforassaystandardsonperformanceofultrasensitiveimmunometricassaysfortshreportofajointwhoifcccollaborativestudy AT brettschneiderh effectsofcommonmatricesforassaystandardsonperformanceofultrasensitiveimmunometricassaysfortshreportofajointwhoifcccollaborativestudy |