Automated synthesis of radiopharmaceuticals for positron emission tomography: an apparatus for [1-(11)C] labeled carboxylic acid

A fully automated apparatus for the synthesis and formulation of short-lived (11)C (t(1/2) = 20 min)-labeled carboxylic acids for positron emission tomograpy (PET) has been developed. Injectable solutions of [1-(11)C]acetic acid, [1-(11)C]octanoic acid and [1-(11)C]palmitic acid wilh radioactivities...

Descripción completa

Detalles Bibliográficos
Autores principales: Yajima, Kazuyoshi, Yamazaki, Hiroyoshi, Kawashima, Hidefumi, Ino, Sento, Hayashi, Nobuyoshi, Miyake, Yoshihiro
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 1995
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2548068/
https://www.ncbi.nlm.nih.gov/pubmed/18925023
http://dx.doi.org/10.1155/S1463924695000198
Descripción
Sumario:A fully automated apparatus for the synthesis and formulation of short-lived (11)C (t(1/2) = 20 min)-labeled carboxylic acids for positron emission tomograpy (PET) has been developed. Injectable solutions of [1-(11)C]acetic acid, [1-(11)C]octanoic acid and [1-(11)C]palmitic acid wilh radioactivities of 6.36-8.29 GBq, 0.070-1.43 GBq and 0.42-0.89 GBq were obtained. The preparation time was under 40 min after the end of bombardment. An automatic washing function means that labeled compound of the same or different kinds can be produced several times a day without any maintenance of the system. The control system is sited away from the ‘hot laboratory’, so operator exposure to radiation is minimized.