Cargando…
Evaluation of a random access analyser: BM/Hitachi 911
The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study f...
Autores principales: | , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
1995
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2548101/ https://www.ncbi.nlm.nih.gov/pubmed/18925048 http://dx.doi.org/10.1155/S1463924695000332 |
_version_ | 1782159402900389888 |
---|---|
author | Kanluan, Taweesook Tangvarasittichai, Surapon Tangvarasittichai, Orathai |
author_facet | Kanluan, Taweesook Tangvarasittichai, Surapon Tangvarasittichai, Orathai |
author_sort | Kanluan, Taweesook |
collection | PubMed |
description | The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtainedfrom BM/Hitachi 911 and 717 were between the extreme values of 0.94 and 1.05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation. |
format | Text |
id | pubmed-2548101 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 1995 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-25481012008-10-16 Evaluation of a random access analyser: BM/Hitachi 911 Kanluan, Taweesook Tangvarasittichai, Surapon Tangvarasittichai, Orathai J Automat Chem Research Article The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtainedfrom BM/Hitachi 911 and 717 were between the extreme values of 0.94 and 1.05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation. Hindawi Publishing Corporation 1995 /pmc/articles/PMC2548101/ /pubmed/18925048 http://dx.doi.org/10.1155/S1463924695000332 Text en Copyright © 1995 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Kanluan, Taweesook Tangvarasittichai, Surapon Tangvarasittichai, Orathai Evaluation of a random access analyser: BM/Hitachi 911 |
title | Evaluation of a random access analyser: BM/Hitachi 911 |
title_full | Evaluation of a random access analyser: BM/Hitachi 911 |
title_fullStr | Evaluation of a random access analyser: BM/Hitachi 911 |
title_full_unstemmed | Evaluation of a random access analyser: BM/Hitachi 911 |
title_short | Evaluation of a random access analyser: BM/Hitachi 911 |
title_sort | evaluation of a random access analyser: bm/hitachi 911 |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2548101/ https://www.ncbi.nlm.nih.gov/pubmed/18925048 http://dx.doi.org/10.1155/S1463924695000332 |
work_keys_str_mv | AT kanluantaweesook evaluationofarandomaccessanalyserbmhitachi911 AT tangvarasittichaisurapon evaluationofarandomaccessanalyserbmhitachi911 AT tangvarasittichaiorathai evaluationofarandomaccessanalyserbmhitachi911 |