Evaluation of a random access analyser: BM/Hitachi 911
The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study f...
Autores principales: | Kanluan, Taweesook, Tangvarasittichai, Surapon, Tangvarasittichai, Orathai |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
1995
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2548101/ https://www.ncbi.nlm.nih.gov/pubmed/18925048 http://dx.doi.org/10.1155/S1463924695000332 |
Ejemplares similares
-
Laboratory evaluation of the COBAS MIRA S random access analyser
por: Kanluan, Taweesook, et al.
Publicado: (1991) -
Multicentre evaluation of the Boehringer Mannheim/Hitachi 911 Analysis System
por: Zaman, Z., et al.
Publicado: (1993) -
Evaluation of the Hitachi 717 analyser
por: Biosca, Carmen, et al.
Publicado: (1989) -
An evaluation of the Hitachi 705 analyser
por: Ferré, C., et al.
Publicado: (1984) -
Multicentre evaluation of the Boehringer Mannheim/Hitachi 917 analysis system
por: van Suijlen, Jeroen D. E., et al.
Publicado: (2000)