A fully automated flow injection atomic absorption system for the determination of copper traces in waters with on-line pre-concentration in an ion-exchange column
The paper describes the development of an automatic on-line column pre-concentration technique using a time based-flow injection atomic absorption spectrometry system. A manifold incorporating a micro-column containing 25 mg of Dowex 50W-X8 was used with a time-based injector for the pre-concentrati...
Autores principales: | , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
1995
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2548107/ https://www.ncbi.nlm.nih.gov/pubmed/18924787 http://dx.doi.org/10.1155/S1463924695000046 |
Sumario: | The paper describes the development of an automatic on-line column pre-concentration technique using a time based-flow injection atomic absorption spectrometry system. A manifold incorporating a micro-column containing 25 mg of Dowex 50W-X8 was used with a time-based injector for the pre-concentration and determination of copper in natural and drinking waters. The system features depend on the alternate positions of a solenoid valve. The 3σ detection limits, enrichment factors, sampling frequency, relative standard deviations and linear calibration graphs were, respectively, in the range 0.6-1.5 μg/l, 25-60, 15-30 measurements/h, 1.0-3.1% and 1-65 μg/ml for pre-concentration times of 1 min. The procedure was successfully applied to a range of water samples and the accuracy was assessed through recovery experiments, the analysis of certified reference water samples and by independent analysis by atomic absorption spectrometry with electrothermal atomization. |
---|