Proposed fibrinogen standard material with purified fibrinogen for plasma fibrinogen measurement on coagulation analyser

A commercially available fibrinogen standard calibrated by a World Health Organization (WHO) reference material is widely used in Japan, and most clinical laboratories use the Clauss method for plasma fibrinogen measurement. However, a current issue in fibrinogen measurement is poor laboratory-to-la...

Descripción completa

Detalles Bibliográficos
Autores principales: Okuda, Masahiro, Uemura, Yahiro, Tatsumi, Noriyuki, Honda, Susumu
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2548388/
https://www.ncbi.nlm.nih.gov/pubmed/18924618
http://dx.doi.org/10.1155/S146392460300018X
Descripción
Sumario:A commercially available fibrinogen standard calibrated by a World Health Organization (WHO) reference material is widely used in Japan, and most clinical laboratories use the Clauss method for plasma fibrinogen measurement. However, a current issue in fibrinogen measurement is poor laboratory-to-laboratory variability. To improve the reliability of fibrinogen values and thereby solve the poor precision and accuracy of plasma fibrinogen testing, the present paper develops a simple and large preparation procedure for a suitable fibrinogen standard and quality control material and evaluates their basic performance. With a new procedure getting high purified fibrinogen by glycine precipitation, the calibrator determined by both the Clauss and Jacobson methods produced a fibrinogen concentration of 2.20 g l(−1). The total precision of the calibrator was excellent (coefficient of variation 1.4-2.1%) in comparison with current plasma fibrinogen materials from the WHO (#98/612) and with a commercial standard (CV 1.9-3.9%). The within-run precision of the calibrator on the coagulation analysers was 1.7-2.8%. Within-analyser variability among the five instruments had good consistency (mean 2.20 ± 0.022 g l(−1); CV 1.0%). The degradation study of the calibrator suggested that storage at 9(°)C for two years was as predicted. In conclusion, the results show that the calibrator prepared herein can be useful as a candidate Japanese fibrinogen standard and is applicable to automated and semi-automated coagulation analysers. Additionally, it is expected that it will be widely used in Japan by diagnostic manufacturers and clinical laboratories as a recommended secondary standard to estimate a fibrinogen value according to the WHO primary standard.