An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system
Recent clinical data indicate that the measurement of the concentration of C-reactive protein (CRP) requires a higher sensitivity and wider dynamic range than most of the current methods can offer. Our goal was to develop a totally automated and highly sensitive CRP assay with an extended range on t...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2000
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2562849/ https://www.ncbi.nlm.nih.gov/pubmed/18924698 http://dx.doi.org/10.1155/S1463924600000195 |
_version_ | 1782159766288596992 |
---|---|
author | Wei, Tie Q. Kramer, Steve Chu, Victor P. Hudson, Dave Kilgore, Daniel Salyer, Sue Parker, Grace Eyberger, Amy Arentzen, Rene Koiv, Heikki |
author_facet | Wei, Tie Q. Kramer, Steve Chu, Victor P. Hudson, Dave Kilgore, Daniel Salyer, Sue Parker, Grace Eyberger, Amy Arentzen, Rene Koiv, Heikki |
author_sort | Wei, Tie Q. |
collection | PubMed |
description | Recent clinical data indicate that the measurement of the concentration of C-reactive protein (CRP) requires a higher sensitivity and wider dynamic range than most of the current methods can offer. Our goal was to develop a totally automated and highly sensitive CRP assay with an extended range on the Dimension(®) clinical chemistry system based on particle-enhanced turbidimetric-immunoassay (PETIA) technology. The improved method was optimized and compared to the Binding Site's radial immunodiffusion assay using disease state specimens to minimize interference. Assay performance was assessed on the Dimension(®) system in a 12-instrument inter-laboratory comparison study. A split-sample comparison (n = 622) was performed between the improved CRP method on the Dimension(®) system and the N Latex CRP mono method on the Behring Nephelometer, using a number of reagent and calibrator lots on multiple instruments. The method was also referenced to the standard material, CRM470, provided by the International Federation of Clinical Chemistry (IFCC). The improved CRP method was linear to 265.1mg/l with a detection limit between 0.2 and 0.5mg/l. The method detects antigen excess from the upper assay limit to 2000mg/l, thereby allowing users to retest the sample with dilution. Calibration was stable for 60 days. The within-run reproducibility (CV) was less than 5.1% and total reproducibility ranged from 1.1 to 6.7% between 3.3 and 265.4mg/l CRP. Linear regression analysis of the results on the improved Dimension(®) method (DM) versus the Behring Nephelometer (BN) yielded the following equation: DM = 0.99 × BN − 0.37; r = 0.992. Minimal interference was observed from sera of patients with elevated IgM, IgG and IgA. The recovery of the IFCC standard was within 100 ± 7 % across multiple lots of reagent and calibrator. The improved CRP method provided a sensitive, accurate and rapid approach to quantify CRP in serum and plasma on the Dimension(®) clinical chemistry system. The ability to detect antigen excess eliminated reporting falsely low results caused by the ‘prozone effect’. |
format | Text |
id | pubmed-2562849 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2000 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-25628492008-10-16 An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system Wei, Tie Q. Kramer, Steve Chu, Victor P. Hudson, Dave Kilgore, Daniel Salyer, Sue Parker, Grace Eyberger, Amy Arentzen, Rene Koiv, Heikki J Autom Methods Manag Chem Research Article Recent clinical data indicate that the measurement of the concentration of C-reactive protein (CRP) requires a higher sensitivity and wider dynamic range than most of the current methods can offer. Our goal was to develop a totally automated and highly sensitive CRP assay with an extended range on the Dimension(®) clinical chemistry system based on particle-enhanced turbidimetric-immunoassay (PETIA) technology. The improved method was optimized and compared to the Binding Site's radial immunodiffusion assay using disease state specimens to minimize interference. Assay performance was assessed on the Dimension(®) system in a 12-instrument inter-laboratory comparison study. A split-sample comparison (n = 622) was performed between the improved CRP method on the Dimension(®) system and the N Latex CRP mono method on the Behring Nephelometer, using a number of reagent and calibrator lots on multiple instruments. The method was also referenced to the standard material, CRM470, provided by the International Federation of Clinical Chemistry (IFCC). The improved CRP method was linear to 265.1mg/l with a detection limit between 0.2 and 0.5mg/l. The method detects antigen excess from the upper assay limit to 2000mg/l, thereby allowing users to retest the sample with dilution. Calibration was stable for 60 days. The within-run reproducibility (CV) was less than 5.1% and total reproducibility ranged from 1.1 to 6.7% between 3.3 and 265.4mg/l CRP. Linear regression analysis of the results on the improved Dimension(®) method (DM) versus the Behring Nephelometer (BN) yielded the following equation: DM = 0.99 × BN − 0.37; r = 0.992. Minimal interference was observed from sera of patients with elevated IgM, IgG and IgA. The recovery of the IFCC standard was within 100 ± 7 % across multiple lots of reagent and calibrator. The improved CRP method provided a sensitive, accurate and rapid approach to quantify CRP in serum and plasma on the Dimension(®) clinical chemistry system. The ability to detect antigen excess eliminated reporting falsely low results caused by the ‘prozone effect’. Hindawi Publishing Corporation 2000 /pmc/articles/PMC2562849/ /pubmed/18924698 http://dx.doi.org/10.1155/S1463924600000195 Text en Copyright © 2000 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Wei, Tie Q. Kramer, Steve Chu, Victor P. Hudson, Dave Kilgore, Daniel Salyer, Sue Parker, Grace Eyberger, Amy Arentzen, Rene Koiv, Heikki An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system |
title | An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system |
title_full | An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system |
title_fullStr | An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system |
title_full_unstemmed | An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system |
title_short | An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system |
title_sort | improved automated immunoassay for c-reactive protein on the dimension(®) clinical chemistry system |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2562849/ https://www.ncbi.nlm.nih.gov/pubmed/18924698 http://dx.doi.org/10.1155/S1463924600000195 |
work_keys_str_mv | AT weitieq animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT kramersteve animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT chuvictorp animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT hudsondave animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT kilgoredaniel animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT salyersue animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT parkergrace animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT eybergeramy animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT arentzenrene animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT koivheikki animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT weitieq improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT kramersteve improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT chuvictorp improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT hudsondave improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT kilgoredaniel improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT salyersue improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT parkergrace improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT eybergeramy improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT arentzenrene improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem AT koivheikki improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem |