An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system

Recent clinical data indicate that the measurement of the concentration of C-reactive protein (CRP) requires a higher sensitivity and wider dynamic range than most of the current methods can offer. Our goal was to develop a totally automated and highly sensitive CRP assay with an extended range on t...

Descripción completa

Detalles Bibliográficos
Autores principales: Wei, Tie Q., Kramer, Steve, Chu, Victor P., Hudson, Dave, Kilgore, Daniel, Salyer, Sue, Parker, Grace, Eyberger, Amy, Arentzen, Rene, Koiv, Heikki
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2000
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2562849/
https://www.ncbi.nlm.nih.gov/pubmed/18924698
http://dx.doi.org/10.1155/S1463924600000195
_version_ 1782159766288596992
author Wei, Tie Q.
Kramer, Steve
Chu, Victor P.
Hudson, Dave
Kilgore, Daniel
Salyer, Sue
Parker, Grace
Eyberger, Amy
Arentzen, Rene
Koiv, Heikki
author_facet Wei, Tie Q.
Kramer, Steve
Chu, Victor P.
Hudson, Dave
Kilgore, Daniel
Salyer, Sue
Parker, Grace
Eyberger, Amy
Arentzen, Rene
Koiv, Heikki
author_sort Wei, Tie Q.
collection PubMed
description Recent clinical data indicate that the measurement of the concentration of C-reactive protein (CRP) requires a higher sensitivity and wider dynamic range than most of the current methods can offer. Our goal was to develop a totally automated and highly sensitive CRP assay with an extended range on the Dimension(®) clinical chemistry system based on particle-enhanced turbidimetric-immunoassay (PETIA) technology. The improved method was optimized and compared to the Binding Site's radial immunodiffusion assay using disease state specimens to minimize interference. Assay performance was assessed on the Dimension(®) system in a 12-instrument inter-laboratory comparison study. A split-sample comparison (n = 622) was performed between the improved CRP method on the Dimension(®) system and the N Latex CRP mono method on the Behring Nephelometer, using a number of reagent and calibrator lots on multiple instruments. The method was also referenced to the standard material, CRM470, provided by the International Federation of Clinical Chemistry (IFCC). The improved CRP method was linear to 265.1mg/l with a detection limit between 0.2 and 0.5mg/l. The method detects antigen excess from the upper assay limit to 2000mg/l, thereby allowing users to retest the sample with dilution. Calibration was stable for 60 days. The within-run reproducibility (CV) was less than 5.1% and total reproducibility ranged from 1.1 to 6.7% between 3.3 and 265.4mg/l CRP. Linear regression analysis of the results on the improved Dimension(®) method (DM) versus the Behring Nephelometer (BN) yielded the following equation: DM = 0.99 × BN − 0.37; r = 0.992. Minimal interference was observed from sera of patients with elevated IgM, IgG and IgA. The recovery of the IFCC standard was within 100 ± 7 % across multiple lots of reagent and calibrator. The improved CRP method provided a sensitive, accurate and rapid approach to quantify CRP in serum and plasma on the Dimension(®) clinical chemistry system. The ability to detect antigen excess eliminated reporting falsely low results caused by the ‘prozone effect’.
format Text
id pubmed-2562849
institution National Center for Biotechnology Information
language English
publishDate 2000
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-25628492008-10-16 An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system Wei, Tie Q. Kramer, Steve Chu, Victor P. Hudson, Dave Kilgore, Daniel Salyer, Sue Parker, Grace Eyberger, Amy Arentzen, Rene Koiv, Heikki J Autom Methods Manag Chem Research Article Recent clinical data indicate that the measurement of the concentration of C-reactive protein (CRP) requires a higher sensitivity and wider dynamic range than most of the current methods can offer. Our goal was to develop a totally automated and highly sensitive CRP assay with an extended range on the Dimension(®) clinical chemistry system based on particle-enhanced turbidimetric-immunoassay (PETIA) technology. The improved method was optimized and compared to the Binding Site's radial immunodiffusion assay using disease state specimens to minimize interference. Assay performance was assessed on the Dimension(®) system in a 12-instrument inter-laboratory comparison study. A split-sample comparison (n = 622) was performed between the improved CRP method on the Dimension(®) system and the N Latex CRP mono method on the Behring Nephelometer, using a number of reagent and calibrator lots on multiple instruments. The method was also referenced to the standard material, CRM470, provided by the International Federation of Clinical Chemistry (IFCC). The improved CRP method was linear to 265.1mg/l with a detection limit between 0.2 and 0.5mg/l. The method detects antigen excess from the upper assay limit to 2000mg/l, thereby allowing users to retest the sample with dilution. Calibration was stable for 60 days. The within-run reproducibility (CV) was less than 5.1% and total reproducibility ranged from 1.1 to 6.7% between 3.3 and 265.4mg/l CRP. Linear regression analysis of the results on the improved Dimension(®) method (DM) versus the Behring Nephelometer (BN) yielded the following equation: DM = 0.99 × BN − 0.37; r = 0.992. Minimal interference was observed from sera of patients with elevated IgM, IgG and IgA. The recovery of the IFCC standard was within 100 ± 7 % across multiple lots of reagent and calibrator. The improved CRP method provided a sensitive, accurate and rapid approach to quantify CRP in serum and plasma on the Dimension(®) clinical chemistry system. The ability to detect antigen excess eliminated reporting falsely low results caused by the ‘prozone effect’. Hindawi Publishing Corporation 2000 /pmc/articles/PMC2562849/ /pubmed/18924698 http://dx.doi.org/10.1155/S1463924600000195 Text en Copyright © 2000 Hindawi Publishing Corporation. http://creativecommons.org/licenses/by/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Wei, Tie Q.
Kramer, Steve
Chu, Victor P.
Hudson, Dave
Kilgore, Daniel
Salyer, Sue
Parker, Grace
Eyberger, Amy
Arentzen, Rene
Koiv, Heikki
An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system
title An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system
title_full An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system
title_fullStr An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system
title_full_unstemmed An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system
title_short An improved automated immunoassay for C-reactive protein on the Dimension(®) clinical chemistry system
title_sort improved automated immunoassay for c-reactive protein on the dimension(®) clinical chemistry system
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2562849/
https://www.ncbi.nlm.nih.gov/pubmed/18924698
http://dx.doi.org/10.1155/S1463924600000195
work_keys_str_mv AT weitieq animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT kramersteve animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT chuvictorp animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT hudsondave animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT kilgoredaniel animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT salyersue animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT parkergrace animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT eybergeramy animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT arentzenrene animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT koivheikki animprovedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT weitieq improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT kramersteve improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT chuvictorp improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT hudsondave improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT kilgoredaniel improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT salyersue improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT parkergrace improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT eybergeramy improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT arentzenrene improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem
AT koivheikki improvedautomatedimmunoassayforcreactiveproteinonthedimensionclinicalchemistrysystem