Monte Carlo simulation for the prediction of precision of absorbance measurements with a miniature CCD spectrometer

The precision characteristics of the absorbance measurements obtained with a low-cost miniature spectrometer incorporating an array detector were evaluated. Uncertainties in absorbance measurements were due to a combination of non-uniform light intensity and detector response over the wavelength ran...

Descripción completa

Detalles Bibliográficos
Autores principales: Tsaousoglou, E. P., Bolis, S. D., Efstathiou, C. E.
Formato: Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2562858/
https://www.ncbi.nlm.nih.gov/pubmed/18924714
http://dx.doi.org/10.1155/S1463924603000063
Descripción
Sumario:The precision characteristics of the absorbance measurements obtained with a low-cost miniature spectrometer incorporating an array detector were evaluated. Uncertainties in absorbance measurements were due to a combination of non-uniform light intensity and detector response over the wavelength range examined (350-850 nm), in conjunction with the digitization of the intensity indications and the intrinsic noise of the detecting elements. The precision characteristics are presented as contour plots displaying the expected RSD% of absorbances on the absorbance versus wavelength plane. The minimum RSD% for the spectrometer configuration tested was observed within the 0.2-1.5 absorbance units and 500-750 nm wavelength range. Without invoking signal enhancement features of the data-acquisition program (scan average, higher integration times, smoothing based on averaging the signal detected by adjacent pixels), the attainable precision within this range was 0.4-0.8%. A computer program based on Monte Carlo simulations was developed for the prediction of absorbance precision characteristics under various conditions of measurements.