Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT
BACKGROUND: Image-guidance systems allow accurate interfractional repositioning of IMRT treatments, however, these may require up to 15 minutes. Therefore intrafraction motion might have an impact on treatment precision. 3D geometric data regarding intrafraction prostate motion are rare; we therefor...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2588616/ https://www.ncbi.nlm.nih.gov/pubmed/18986517 http://dx.doi.org/10.1186/1748-717X-3-37 |
_version_ | 1782160965183209472 |
---|---|
author | Boda-Heggemann, Judit Köhler, Frederick Marc Wertz, Hansjörg Ehmann, Michael Hermann, Brigitte Riesenacker, Nadja Küpper, Beate Lohr, Frank Wenz, Frederik |
author_facet | Boda-Heggemann, Judit Köhler, Frederick Marc Wertz, Hansjörg Ehmann, Michael Hermann, Brigitte Riesenacker, Nadja Küpper, Beate Lohr, Frank Wenz, Frederik |
author_sort | Boda-Heggemann, Judit |
collection | PubMed |
description | BACKGROUND: Image-guidance systems allow accurate interfractional repositioning of IMRT treatments, however, these may require up to 15 minutes. Therefore intrafraction motion might have an impact on treatment precision. 3D geometric data regarding intrafraction prostate motion are rare; we therefore assessed its magnitude with pre- and post-treatment fiducial-based imaging with cone-beam-CT (CBCT). METHODS: 39 IMRT fractions in 5 prostate cancer patients after (125)I-seed implantation were evaluated. Patient position was corrected based on the (125)I-seeds after pre-treatment CBCT. Immediately after treatment delivery, a second CBCT was performed. Differences in bone- and fiducial position were measured by seed-based grey-value matching. RESULTS: Fraction time was 13.6 ± 1.6 minutes. Median overall displacement vector length of (125)I-seeds was 3 mm (M = 3 mm, Σ = 0.9 mm, σ = 1.7 mm; M: group systematic error, Σ: SD of systematic error, σ: SD of random error). Median displacement vector of bony structures was 1.84 mm (M = 2.9 mm, Σ = 1 mm, σ = 3.2 mm). Median displacement vector length of the prostate relative to bony structures was 1.9 mm (M = 3 mm, Σ = 1.3 mm, σ = 2.6 mm). CONCLUSION: a) Overall displacement vector length during an IMRT session is < 3 mm. b) Positioning devices reducing intrafraction bony displacements can further reduce overall intrafraction motion. c) Intrafraction prostate motion relative to bony structures is < 2 mm and may be further reduced by institutional protocols and reduction of IMRT duration. |
format | Text |
id | pubmed-2588616 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-25886162008-11-28 Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT Boda-Heggemann, Judit Köhler, Frederick Marc Wertz, Hansjörg Ehmann, Michael Hermann, Brigitte Riesenacker, Nadja Küpper, Beate Lohr, Frank Wenz, Frederik Radiat Oncol Research BACKGROUND: Image-guidance systems allow accurate interfractional repositioning of IMRT treatments, however, these may require up to 15 minutes. Therefore intrafraction motion might have an impact on treatment precision. 3D geometric data regarding intrafraction prostate motion are rare; we therefore assessed its magnitude with pre- and post-treatment fiducial-based imaging with cone-beam-CT (CBCT). METHODS: 39 IMRT fractions in 5 prostate cancer patients after (125)I-seed implantation were evaluated. Patient position was corrected based on the (125)I-seeds after pre-treatment CBCT. Immediately after treatment delivery, a second CBCT was performed. Differences in bone- and fiducial position were measured by seed-based grey-value matching. RESULTS: Fraction time was 13.6 ± 1.6 minutes. Median overall displacement vector length of (125)I-seeds was 3 mm (M = 3 mm, Σ = 0.9 mm, σ = 1.7 mm; M: group systematic error, Σ: SD of systematic error, σ: SD of random error). Median displacement vector of bony structures was 1.84 mm (M = 2.9 mm, Σ = 1 mm, σ = 3.2 mm). Median displacement vector length of the prostate relative to bony structures was 1.9 mm (M = 3 mm, Σ = 1.3 mm, σ = 2.6 mm). CONCLUSION: a) Overall displacement vector length during an IMRT session is < 3 mm. b) Positioning devices reducing intrafraction bony displacements can further reduce overall intrafraction motion. c) Intrafraction prostate motion relative to bony structures is < 2 mm and may be further reduced by institutional protocols and reduction of IMRT duration. BioMed Central 2008-11-05 /pmc/articles/PMC2588616/ /pubmed/18986517 http://dx.doi.org/10.1186/1748-717X-3-37 Text en Copyright © 2008 Boda-Heggemann et al; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Boda-Heggemann, Judit Köhler, Frederick Marc Wertz, Hansjörg Ehmann, Michael Hermann, Brigitte Riesenacker, Nadja Küpper, Beate Lohr, Frank Wenz, Frederik Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT |
title | Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT |
title_full | Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT |
title_fullStr | Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT |
title_full_unstemmed | Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT |
title_short | Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT |
title_sort | intrafraction motion of the prostate during an imrt session: a fiducial-based 3d measurement with cone-beam ct |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2588616/ https://www.ncbi.nlm.nih.gov/pubmed/18986517 http://dx.doi.org/10.1186/1748-717X-3-37 |
work_keys_str_mv | AT bodaheggemannjudit intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT kohlerfrederickmarc intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT wertzhansjorg intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT ehmannmichael intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT hermannbrigitte intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT riesenackernadja intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT kupperbeate intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT lohrfrank intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct AT wenzfrederik intrafractionmotionoftheprostateduringanimrtsessionafiducialbased3dmeasurementwithconebeamct |