Cargando…

Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing

The clinical efficacy of ICD-CRT therapy depends on accurate sensing of intracardiac signals and sensing algorithms. We report the occurrence of sensing abnormality in a patient with ICD-CRT. In this patient, oversensing of myopotentials during strenuous muscular activity resulted in an inappropriat...

Descripción completa

Detalles Bibliográficos
Autores principales: Barakpour, Hamid, Emkanjoo, Zahra, Alizadeh, Abolfath, Sadr-Ameli, Mohammad Ali
Formato: Texto
Lenguaje:English
Publicado: Indian Heart Rhythm Society 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2615066/
https://www.ncbi.nlm.nih.gov/pubmed/19165363
_version_ 1782163283553288192
author Barakpour, Hamid
Emkanjoo, Zahra
Alizadeh, Abolfath
Sadr-Ameli, Mohammad Ali
author_facet Barakpour, Hamid
Emkanjoo, Zahra
Alizadeh, Abolfath
Sadr-Ameli, Mohammad Ali
author_sort Barakpour, Hamid
collection PubMed
description The clinical efficacy of ICD-CRT therapy depends on accurate sensing of intracardiac signals and sensing algorithms. We report the occurrence of sensing abnormality in a patient with ICD-CRT. In this patient, oversensing of myopotentials during strenuous muscular activity resulted in an inappropriate ICD-CRT discharge. Although modern ICDs are highly effective in detecting and terminating malignant tachyarrhythmias, their detection specificity must be improved. It is possible to find the mechanism of arrhythmia by EGM. Simple device reprogramming make it possible to avoid the oversensing of myopotentials.
format Text
id pubmed-2615066
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher Indian Heart Rhythm Society
record_format MEDLINE/PubMed
spelling pubmed-26150662009-01-22 Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing Barakpour, Hamid Emkanjoo, Zahra Alizadeh, Abolfath Sadr-Ameli, Mohammad Ali Indian Pacing Electrophysiol J Case Report The clinical efficacy of ICD-CRT therapy depends on accurate sensing of intracardiac signals and sensing algorithms. We report the occurrence of sensing abnormality in a patient with ICD-CRT. In this patient, oversensing of myopotentials during strenuous muscular activity resulted in an inappropriate ICD-CRT discharge. Although modern ICDs are highly effective in detecting and terminating malignant tachyarrhythmias, their detection specificity must be improved. It is possible to find the mechanism of arrhythmia by EGM. Simple device reprogramming make it possible to avoid the oversensing of myopotentials. Indian Heart Rhythm Society 2009-01-07 /pmc/articles/PMC2615066/ /pubmed/19165363 Text en Copyright: © 2009 Barakpour et al. http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Barakpour, Hamid
Emkanjoo, Zahra
Alizadeh, Abolfath
Sadr-Ameli, Mohammad Ali
Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing
title Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing
title_full Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing
title_fullStr Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing
title_full_unstemmed Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing
title_short Inappropriate Shock Delivered By Implantable Cardioverter Defibrillator - Cardiac Resynchronization Therapy (ICD-CRT) Due To Myopotential Oversensing
title_sort inappropriate shock delivered by implantable cardioverter defibrillator - cardiac resynchronization therapy (icd-crt) due to myopotential oversensing
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2615066/
https://www.ncbi.nlm.nih.gov/pubmed/19165363
work_keys_str_mv AT barakpourhamid inappropriateshockdeliveredbyimplantablecardioverterdefibrillatorcardiacresynchronizationtherapyicdcrtduetomyopotentialoversensing
AT emkanjoozahra inappropriateshockdeliveredbyimplantablecardioverterdefibrillatorcardiacresynchronizationtherapyicdcrtduetomyopotentialoversensing
AT alizadehabolfath inappropriateshockdeliveredbyimplantablecardioverterdefibrillatorcardiacresynchronizationtherapyicdcrtduetomyopotentialoversensing
AT sadramelimohammadali inappropriateshockdeliveredbyimplantablecardioverterdefibrillatorcardiacresynchronizationtherapyicdcrtduetomyopotentialoversensing