Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway
OBJECTIVE: To assess the influence of variable factors such as the size of the airway and the CT imaging parameters such as the reconstruction kernel, field-of-view (FOV), and slice thickness on the automatic measurement of airway dimension. MATERIALS AND METHODS: An airway phantom was fabricated th...
Autores principales: | , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
The Korean Radiological Society
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2627250/ https://www.ncbi.nlm.nih.gov/pubmed/18525225 http://dx.doi.org/10.3348/kjr.2008.9.3.226 |
_version_ | 1782163526342672384 |
---|---|
author | Kim, Namkug Seo, Joon Beom Song, Koun-Sik Chae, Eun Jin Kang, Suk-Ho |
author_facet | Kim, Namkug Seo, Joon Beom Song, Koun-Sik Chae, Eun Jin Kang, Suk-Ho |
author_sort | Kim, Namkug |
collection | PubMed |
description | OBJECTIVE: To assess the influence of variable factors such as the size of the airway and the CT imaging parameters such as the reconstruction kernel, field-of-view (FOV), and slice thickness on the automatic measurement of airway dimension. MATERIALS AND METHODS: An airway phantom was fabricated that contained eleven poly-acryl tubes of various lumen diameters and wall thicknesses. The measured density of the poly-acryl wall was 150 HU, and the measured density of the airspace filled with polyurethane foam was -900 HU. CT images were obtained using a 16-MDCT (multidetector CT) scanner and were reconstructed with various reconstruction kernels, thicknesses and FOV. The luminal radius and wall thickness were measured using in-house software based on the full-width-half-maximum method. The measured values as determined by CT and the actual dimensions of the tubes were compared. RESULTS: Measurements were most accurate on images reconstructed with use of a standard kernel (mean error: -0.03 ± 0.21 mm for wall thickness and -0.12 ± 0.11 mm for the luminal radius). There was no significant difference in accuracy among images with the use of variable slice thicknesses or a variable FOV. Below a 1-mm threshold, the measurement failed to represent the change of the real dimensions. CONCLUSION: Measurement accuracy was strongly influenced by the specific reconstruction kernel utilized. For accurate measurement, standardization of the imaging protocol and selection of the appropriate anatomic level are essential. |
format | Text |
id | pubmed-2627250 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | The Korean Radiological Society |
record_format | MEDLINE/PubMed |
spelling | pubmed-26272502009-02-17 Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway Kim, Namkug Seo, Joon Beom Song, Koun-Sik Chae, Eun Jin Kang, Suk-Ho Korean J Radiol Original Article OBJECTIVE: To assess the influence of variable factors such as the size of the airway and the CT imaging parameters such as the reconstruction kernel, field-of-view (FOV), and slice thickness on the automatic measurement of airway dimension. MATERIALS AND METHODS: An airway phantom was fabricated that contained eleven poly-acryl tubes of various lumen diameters and wall thicknesses. The measured density of the poly-acryl wall was 150 HU, and the measured density of the airspace filled with polyurethane foam was -900 HU. CT images were obtained using a 16-MDCT (multidetector CT) scanner and were reconstructed with various reconstruction kernels, thicknesses and FOV. The luminal radius and wall thickness were measured using in-house software based on the full-width-half-maximum method. The measured values as determined by CT and the actual dimensions of the tubes were compared. RESULTS: Measurements were most accurate on images reconstructed with use of a standard kernel (mean error: -0.03 ± 0.21 mm for wall thickness and -0.12 ± 0.11 mm for the luminal radius). There was no significant difference in accuracy among images with the use of variable slice thicknesses or a variable FOV. Below a 1-mm threshold, the measurement failed to represent the change of the real dimensions. CONCLUSION: Measurement accuracy was strongly influenced by the specific reconstruction kernel utilized. For accurate measurement, standardization of the imaging protocol and selection of the appropriate anatomic level are essential. The Korean Radiological Society 2008 2008-06-20 /pmc/articles/PMC2627250/ /pubmed/18525225 http://dx.doi.org/10.3348/kjr.2008.9.3.226 Text en Copyright © 2008 The Korean Radiological Society http://creativecommons.org/licenses/by-nc/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Kim, Namkug Seo, Joon Beom Song, Koun-Sik Chae, Eun Jin Kang, Suk-Ho Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway |
title | Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway |
title_full | Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway |
title_fullStr | Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway |
title_full_unstemmed | Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway |
title_short | Semi-Automatic Measurement of the Airway Dimension by Computed Tomography Using the Full-Width-Half-Maximum Method: a Study on the Measurement Accuracy according to the CT Parameters and Size of the Airway |
title_sort | semi-automatic measurement of the airway dimension by computed tomography using the full-width-half-maximum method: a study on the measurement accuracy according to the ct parameters and size of the airway |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2627250/ https://www.ncbi.nlm.nih.gov/pubmed/18525225 http://dx.doi.org/10.3348/kjr.2008.9.3.226 |
work_keys_str_mv | AT kimnamkug semiautomaticmeasurementoftheairwaydimensionbycomputedtomographyusingthefullwidthhalfmaximummethodastudyonthemeasurementaccuracyaccordingtothectparametersandsizeoftheairway AT seojoonbeom semiautomaticmeasurementoftheairwaydimensionbycomputedtomographyusingthefullwidthhalfmaximummethodastudyonthemeasurementaccuracyaccordingtothectparametersandsizeoftheairway AT songkounsik semiautomaticmeasurementoftheairwaydimensionbycomputedtomographyusingthefullwidthhalfmaximummethodastudyonthemeasurementaccuracyaccordingtothectparametersandsizeoftheairway AT chaeeunjin semiautomaticmeasurementoftheairwaydimensionbycomputedtomographyusingthefullwidthhalfmaximummethodastudyonthemeasurementaccuracyaccordingtothectparametersandsizeoftheairway AT kangsukho semiautomaticmeasurementoftheairwaydimensionbycomputedtomographyusingthefullwidthhalfmaximummethodastudyonthemeasurementaccuracyaccordingtothectparametersandsizeoftheairway |