Cargando…
Tremor suppression in ECG
BACKGROUND: Electrocardiogram recordings are very often contaminated by high-frequency noise usually power-line interference and EMG disturbances (tremor). Specific method for interference cancellation without affecting the proper ECG components, called subtraction procedure, was developed some two...
Autores principales: | , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2645399/ https://www.ncbi.nlm.nih.gov/pubmed/19019218 http://dx.doi.org/10.1186/1475-925X-7-29 |
_version_ | 1782164781650673664 |
---|---|
author | Dotsinsky, Ivan A Mihov, Georgy S |
author_facet | Dotsinsky, Ivan A Mihov, Georgy S |
author_sort | Dotsinsky, Ivan A |
collection | PubMed |
description | BACKGROUND: Electrocardiogram recordings are very often contaminated by high-frequency noise usually power-line interference and EMG disturbances (tremor). Specific method for interference cancellation without affecting the proper ECG components, called subtraction procedure, was developed some two decades ago. Filtering out the tremor remains a priori partially successful since it has a relatively wide spectrum, which overlaps the useful ECG frequency band. METHOD: The proposed method for tremor suppression implements the following three procedures. Contaminated ECG signals are subjected to moving averaging (comb filter with linear phase characteristic) with first zero set at 50 Hz to suppress tremor and PL interference simultaneously. The reduced peaks of QRS complexes and other relatively high and steep ECG waves are then restored by an introduced by us procedure called linearly-angular, so that the useful high frequency components are preserved in the range specified by the embedded in the ECG instrument filter, usually up to 125 Hz. Finally, a Savitzky-Golay smoothing filter is applied for supplementary tremor suppression outside the QRS complexes. RESULTS: The results obtained show a low level of the residual EMG disturbances together with negligible distortion of the wave shapes regardless of rhythm and morphology changes. |
format | Text |
id | pubmed-2645399 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-26453992009-02-20 Tremor suppression in ECG Dotsinsky, Ivan A Mihov, Georgy S Biomed Eng Online Research BACKGROUND: Electrocardiogram recordings are very often contaminated by high-frequency noise usually power-line interference and EMG disturbances (tremor). Specific method for interference cancellation without affecting the proper ECG components, called subtraction procedure, was developed some two decades ago. Filtering out the tremor remains a priori partially successful since it has a relatively wide spectrum, which overlaps the useful ECG frequency band. METHOD: The proposed method for tremor suppression implements the following three procedures. Contaminated ECG signals are subjected to moving averaging (comb filter with linear phase characteristic) with first zero set at 50 Hz to suppress tremor and PL interference simultaneously. The reduced peaks of QRS complexes and other relatively high and steep ECG waves are then restored by an introduced by us procedure called linearly-angular, so that the useful high frequency components are preserved in the range specified by the embedded in the ECG instrument filter, usually up to 125 Hz. Finally, a Savitzky-Golay smoothing filter is applied for supplementary tremor suppression outside the QRS complexes. RESULTS: The results obtained show a low level of the residual EMG disturbances together with negligible distortion of the wave shapes regardless of rhythm and morphology changes. BioMed Central 2008-11-19 /pmc/articles/PMC2645399/ /pubmed/19019218 http://dx.doi.org/10.1186/1475-925X-7-29 Text en Copyright © 2008 Dotsinsky and Mihov; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( (http://creativecommons.org/licenses/by/2.0) ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Dotsinsky, Ivan A Mihov, Georgy S Tremor suppression in ECG |
title | Tremor suppression in ECG |
title_full | Tremor suppression in ECG |
title_fullStr | Tremor suppression in ECG |
title_full_unstemmed | Tremor suppression in ECG |
title_short | Tremor suppression in ECG |
title_sort | tremor suppression in ecg |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2645399/ https://www.ncbi.nlm.nih.gov/pubmed/19019218 http://dx.doi.org/10.1186/1475-925X-7-29 |
work_keys_str_mv | AT dotsinskyivana tremorsuppressioninecg AT mihovgeorgys tremorsuppressioninecg |