Cargando…

Determination of interfacial fracture toughness of bone–cement interface using sandwich Brazilian disks

The long-term stability of cemented total hip replacements critically depends on the lasting integrity of the bond between bone and bone cement. Conventionally, the bonding strength of bone–cement is obtained by mechanical tests that tend to produce a large variability between specimens and test met...

Descripción completa

Detalles Bibliográficos
Autores principales: Tong, J., Wong, K.Y., Lupton, C.
Formato: Texto
Lenguaje:English
Publicado: Pergamon Press 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2661062/
https://www.ncbi.nlm.nih.gov/pubmed/19330043
http://dx.doi.org/10.1016/j.engfracmech.2006.02.014
Descripción
Sumario:The long-term stability of cemented total hip replacements critically depends on the lasting integrity of the bond between bone and bone cement. Conventionally, the bonding strength of bone–cement is obtained by mechanical tests that tend to produce a large variability between specimens and test methods. In this work, interfacial fracture toughness of synthetic bone–cement interface has been studied using sandwiched Brazilian disk specimens. Experiments were carried out using polyurethane foams as substrates and a common bone cement as an interlayer. Selected loading angles from 0° to 25° were used to achieve full loading conditions from mode I to mode II. Finite element analyses were carried out to obtain the solutions for strain energy release rates at given phase angles associated with the experimental models. The effects of crack length on the measured interfacial fracture toughness were examined. Microscopic studies were also carried out to obtain the morphology of the fractured interfaces at selected loading angles. The implication of the results on the assessment of fixation in acetabular replacements is discussed in the light of preliminary work on bovine cancellous bone–cement interface.