Correlation Between Intrasac Pressure Measurements of a Pressure Sensor and an Angiographic Catheter During Endovascular Repair of Abdominal Aortic Aneurysm

PURPOSE: To establish a correlation between intrasac pressure measurements of a pressure sensor and an angiographic catheter placed in the same aneurysm sac before and after its exclusion by an endoprosthesis. METHODS: Patients who underwent endovascular abdominal aortic aneurysm repair and received...

Descripción completa

Detalles Bibliográficos
Autores principales: Silveira, Pierre Galvagni, Miller, Christopher William Teixeira, Mendes, Rafael Freygang, Galego, Gilberto Nascimento
Formato: Texto
Lenguaje:English
Publicado: Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2664183/
https://www.ncbi.nlm.nih.gov/pubmed/18297208
Descripción
Sumario:PURPOSE: To establish a correlation between intrasac pressure measurements of a pressure sensor and an angiographic catheter placed in the same aneurysm sac before and after its exclusion by an endoprosthesis. METHODS: Patients who underwent endovascular abdominal aortic aneurysm repair and received an EndoSure(TM) wireless pressure sensor implant between March 19 and December 11, 2004 were enrolled in the study. Simultaneous readings of systolic, diastolic, mean, and pulse pressure within the aneurysm sac were obtained from the catheter and the sensor, both before and after sac exclusion by the endoprosthesis (Readings 1 and 2, respectively). Intrasac pressure measurements were compared using Pearson’s correlation and Student’s t test. Statistical significance was set at p<0.05. RESULTS: Twenty-five patients had the pressure sensor implanted, with simultaneous readings (i.e., recorded by both devices) obtained in 19 patients for Reading 1 and in 10 patients for Reading 2. There was a statistically significant correlation for all pressure variables during both readings, with p<0.01 for all except the pulse pressure in Reading 1 (p<0.05). Statistical significance of pressure variations before and after abdominal aortic aneurysm exclusion was coincident between the sensor and catheter for diastolic (p>0.05), mean (p>0.05), and pulse (p<0.01) pressures; the sole disagreement was observed for systolic pressure, which varied, on average, 31.23 mmHg by the catheter (p<0.05) and 22 mmHg (p>0.05) by the sensor. CONCLUSION: The excellent agreement between intrasac pressure readings recorded by the catheter and the sensor justifies use of the latter for detection of post-exclusion abdominal aortic aneurysm pressurization.