Reproducibility of Regional Pulse Wave Velocity in Healthy Subjects

BACKGROUND/AIMS: Despite the clinical importance and widespread use of pulse wave velocity (PWV), there are no standards for pulse sensors or for system requirements to ensure accurate pulse wave measurement. We assessed the reproducibility of PWV values using a newly developed PWV measurement syste...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Nak Bum, Park, Chang Gyu
Formato: Texto
Lenguaje:English
Publicado: The Korean Association of Internal Medicine 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2687650/
https://www.ncbi.nlm.nih.gov/pubmed/19270477
http://dx.doi.org/10.3904/kjim.2009.24.1.19
Descripción
Sumario:BACKGROUND/AIMS: Despite the clinical importance and widespread use of pulse wave velocity (PWV), there are no standards for pulse sensors or for system requirements to ensure accurate pulse wave measurement. We assessed the reproducibility of PWV values using a newly developed PWV measurement system. METHODS: The system used in this study was the PP-1000, which simultaneously provides regional PWV values from arteries at four different sites (carotid, femoral, radial, and dorsalis pedis). Seventeen healthy male subjects without any cardiovascular disease participated in this study. Two observers performed two consecutive measurements in the same subject in random order. To evaluate the reproducibility of the system, two sets of analyses (within-observer and between-observer) were performed. RESULTS: The means±SD of PWV for the aorta, arm, and leg were 7.0±1.48, 8.43±1.14, and 8.09±0.98 m/s as measured by observer A and 6.76±1.00, 7.97±0.80, and 7.97±0.72 m/s by observer B, respectively. Between-observer differences for the aorta, arm, and leg were 0.14±0.62, 0.18±0.84, and 0.07±0.86 m/s, respectively, and the correlation coefficients were high, especially for aortic PWV (r=0.93). All the measurements showed significant correlation coefficients, ranging from 0.94 to 0.99. CONCLUSIONS: The PWV measurement system used in this study provides accurate analysis results with high reproducibility. It is necessary to provide an accurate algorithm for the detection of additional features such as flow wave, reflection wave, and dicrotic notch from a pulse waveform.