Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences
Serum parameters as indicators for the efficacy of therapeutic drugs are currently in the focus of intensive research. The induction of certain cytokines (or cytokine patterns) is known to be related to the status of the immune response e.g. in regulating the T(H)1/T(H)2 balance. Regarding their pot...
Autores principales: | , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Libertas Academica
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2688352/ https://www.ncbi.nlm.nih.gov/pubmed/19578507 |
_version_ | 1782167697009672192 |
---|---|
author | Nechansky, Andreas Grunt, Susanne Roitt, Ivan M. Kircheis, Ralf |
author_facet | Nechansky, Andreas Grunt, Susanne Roitt, Ivan M. Kircheis, Ralf |
author_sort | Nechansky, Andreas |
collection | PubMed |
description | Serum parameters as indicators for the efficacy of therapeutic drugs are currently in the focus of intensive research. The induction of certain cytokines (or cytokine patterns) is known to be related to the status of the immune response e.g. in regulating the T(H)1/T(H)2 balance. Regarding their potential value as surrogate parameters in clinical trials and subsequently for the assignment of treatment efficacy, the accurate and reliable determination of cytokines in patient serum is mandatory. Because serum samples are precious and limited, test methods—like the xMAP multiplex technology—that allow for the simultaneous determination of a variety of cytokines from only a small sample aliquot, can offer great advantages. We here have compared multiplex kits from three different manufactures and found striking differences upon standardizing using WHO standards for selected cytokines. We therefore extended our xMAP multiplex measurements investigations to an ex-vivo situation by testing serum samples and found that the cytokine amounts measured was critically influenced by the actual kit used. The presented data indicate that statements regarding the quantitive determination of cytokines—and therefore their use as biomarkers—in serum samples have to be interpreted with caution. |
format | Text |
id | pubmed-2688352 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | Libertas Academica |
record_format | MEDLINE/PubMed |
spelling | pubmed-26883522009-07-01 Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences Nechansky, Andreas Grunt, Susanne Roitt, Ivan M. Kircheis, Ralf Biomark Insights Original Research Serum parameters as indicators for the efficacy of therapeutic drugs are currently in the focus of intensive research. The induction of certain cytokines (or cytokine patterns) is known to be related to the status of the immune response e.g. in regulating the T(H)1/T(H)2 balance. Regarding their potential value as surrogate parameters in clinical trials and subsequently for the assignment of treatment efficacy, the accurate and reliable determination of cytokines in patient serum is mandatory. Because serum samples are precious and limited, test methods—like the xMAP multiplex technology—that allow for the simultaneous determination of a variety of cytokines from only a small sample aliquot, can offer great advantages. We here have compared multiplex kits from three different manufactures and found striking differences upon standardizing using WHO standards for selected cytokines. We therefore extended our xMAP multiplex measurements investigations to an ex-vivo situation by testing serum samples and found that the cytokine amounts measured was critically influenced by the actual kit used. The presented data indicate that statements regarding the quantitive determination of cytokines—and therefore their use as biomarkers—in serum samples have to be interpreted with caution. Libertas Academica 2008-04-18 /pmc/articles/PMC2688352/ /pubmed/19578507 Text en © 2008 by the authors http://creativecommons.org/licenses/by/3.0 This article is published under the Creative Commons Attribution By licence. For further information go to: http://creativecommons.org/licenses/by/3.0. (http://creativecommons.org/licenses/by/3.0) |
spellingShingle | Original Research Nechansky, Andreas Grunt, Susanne Roitt, Ivan M. Kircheis, Ralf Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences |
title | Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences |
title_full | Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences |
title_fullStr | Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences |
title_full_unstemmed | Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences |
title_short | Comparison of the Calibration Standards of Three Commercially Available Multiplex Kits for Human Cytokine Measurement to WHO Standards Reveals Striking Differences |
title_sort | comparison of the calibration standards of three commercially available multiplex kits for human cytokine measurement to who standards reveals striking differences |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2688352/ https://www.ncbi.nlm.nih.gov/pubmed/19578507 |
work_keys_str_mv | AT nechanskyandreas comparisonofthecalibrationstandardsofthreecommerciallyavailablemultiplexkitsforhumancytokinemeasurementtowhostandardsrevealsstrikingdifferences AT gruntsusanne comparisonofthecalibrationstandardsofthreecommerciallyavailablemultiplexkitsforhumancytokinemeasurementtowhostandardsrevealsstrikingdifferences AT roittivanm comparisonofthecalibrationstandardsofthreecommerciallyavailablemultiplexkitsforhumancytokinemeasurementtowhostandardsrevealsstrikingdifferences AT kircheisralf comparisonofthecalibrationstandardsofthreecommerciallyavailablemultiplexkitsforhumancytokinemeasurementtowhostandardsrevealsstrikingdifferences |