Cargando…

Long-Term Safety and Effectiveness of Style 410 Highly Cohesive Silicone Breast Implants

BACKGROUND: In 2006, a single-center Swedish study demonstrated a low rupture rate and high patient satisfaction with the Style 410 shaped, form-stable gel implant. The current study aimed to validate the accuracy of the previously published results across multiple European sites. METHODS: A total o...

Descripción completa

Detalles Bibliográficos
Autores principales: Hedén, Per, Bronz, Giorgio, Elberg, Jens Jorgen, Deraemaecker, Rika, Murphy, Diane K., Slicton, Araceli, Brenner, R. James, Svarvar, Catarina, van Tetering, Joost, van der Weij, Leo Peter
Formato: Texto
Lenguaje:English
Publicado: Springer-Verlag 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2693778/
https://www.ncbi.nlm.nih.gov/pubmed/19437071
http://dx.doi.org/10.1007/s00266-009-9360-x
Descripción
Sumario:BACKGROUND: In 2006, a single-center Swedish study demonstrated a low rupture rate and high patient satisfaction with the Style 410 shaped, form-stable gel implant. The current study aimed to validate the accuracy of the previously published results across multiple European sites. METHODS: A total of 163 subjects (~70% had augmentation [n = 112], 15% had reconstruction [n = 25], and 15% had revision [n = 26]) underwent a physical examination followed by breast magnetic resonance imaging (MRI) for rupture detection. These subjects had been implanted for 5 to 11 years with at least one Style 410 shaped gel breast implant before examination. The secondary end points included lactation, reproductive and breast disease history before and after implantation, and quality-of-life measurements and complications after implantation. RESULTS: The implant rupture rate was 1.7% a median of 8 years after implantation. Capsular contracture was the most common complication noted at the physical examination, occurring for 5.3% of implants, and there were no cases of grade 4 capsular contracture. The postimplantation rates for lactation and reproductive problems and breast disease were lower than the preimplantation rates. Breast implantation surgery was considered advantageous by 91% of the subjects, demonstrating high patient satisfaction. CONCLUSIONS: The Style 410 anatomically shaped, form-stable gel breast implants demonstrated long-term safety and effectiveness.