Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers
OBJECTIVE: To determine the optimized protocol for wet monopolar radiofrequency ablation (RFA) using a perfused-cooled electrode to induce coagulation necrosis in the ex vivo bovine liver. MATERIALS AND METHODS: Radiofrequency was applied to excised bovine livers in a monopolar mode using a 200W gen...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
The Korean Radiological Society
2004
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2698169/ https://www.ncbi.nlm.nih.gov/pubmed/15637475 http://dx.doi.org/10.3348/kjr.2004.5.4.250 |
_version_ | 1782168387715072000 |
---|---|
author | Lee, Jeong Min Han, Joon Koo Kim, Se Hyung Lee, Jae Young Shin, Kyung Sook Han, Chang Jin Lee, Min Woo Choi, Jun Il Choi, Byung Ihn |
author_facet | Lee, Jeong Min Han, Joon Koo Kim, Se Hyung Lee, Jae Young Shin, Kyung Sook Han, Chang Jin Lee, Min Woo Choi, Jun Il Choi, Byung Ihn |
author_sort | Lee, Jeong Min |
collection | PubMed |
description | OBJECTIVE: To determine the optimized protocol for wet monopolar radiofrequency ablation (RFA) using a perfused-cooled electrode to induce coagulation necrosis in the ex vivo bovine liver. MATERIALS AND METHODS: Radiofrequency was applied to excised bovine livers in a monopolar mode using a 200W generator with an internally cooled electrode (groups A and B) or a perfused-cooled electrode (groups C, D, E, and F) at maximum power (150-200 W) for 10 minutes. A total of 60 ablation zones were created with six different regimens: group A - dry RFA using intra-electrode cooling; group B - dry RFA using intra-electrode cooling and a pulsing algorithm; group C - wet RFA using only interstitial hypertonic saline (HS) infusion; group D - wet RFA using interstitial HS infusion and a pulsing algorithm; group E - wet RFA using interstitial HS infusion and intra-electrode cooling; and group F - wet RFA using interstitial HS infusion, intra-electrode cooling and a pulsing algorithm. In groups C, D, E, and F, RFA was performed with the infusion of 6% HS through the perfused cooled electrode at a rate of 2 mL/minute. During RFA, we measured the tissue temperature at a distance of 15 mm from the electrode. The dimensions of the ablation zones and the changes in impedance, currents, and liver temperature during RFA were compared between these six groups. RESULTS: During RFA, the mean tissue impedances in groups A (243 ± 88 Ω) and C (252.5 ± 108 Ω) were significantly higher than those in groups B (85 ± 18.7 Ω), D (108.2 ± 85 Ω), E (70.0 ± 16.3 Ω), and F (66.5 ± 7 Ω) (p < 0.05). The mean currents in groups E and F were significantly higher than those in groups B and D, which were significantly higher than those in groups A and C (p < 0.05): 520 ± 425 mA in group A, 1163 ± 34 mA in group B, 652.5 ± 418 mA in group C, 842.5 ± 773 mA in group D, 1665 ± 295 mA in group E, and 1830 ± 109 mA in group F. The mean volumes of the ablation regions in groups E and F were significantly larger than those in the other groups (p < 0.05): 17.7 ± 5.6 cm(3) in group A, 34.5 ± 3.0 cm(3) in group B, 20.2 ± 15.6 cm(3) in group C, 36.1 ± 19.5 cm(3) in group D, 68.1 ± 12.4 cm(3) in group E, and 79.5 ± 31 cm(3) in group F. The final tissue temperatures at a distance of 15 mm from the electrode were higher in groups E and F than those in groups A, C, and D (p < 0.05): 50 ± 7.5℃ in group A, 66 ± 13.6℃ in group B, 60 ± 13.4℃ in group C, 61 ±12.7℃ in group D, 78 ± 14.2℃ in group E, and 79 ± 12.0℃ in group F. CONCLUSION: Wet monopolar RFA, using intra-electrode cooling and interstitial saline infusion, showed better performance in creating a large ablation zone than either dry RFA or wet RFA without intra-electrode cooling. |
format | Text |
id | pubmed-2698169 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2004 |
publisher | The Korean Radiological Society |
record_format | MEDLINE/PubMed |
spelling | pubmed-26981692009-06-23 Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers Lee, Jeong Min Han, Joon Koo Kim, Se Hyung Lee, Jae Young Shin, Kyung Sook Han, Chang Jin Lee, Min Woo Choi, Jun Il Choi, Byung Ihn Korean J Radiol Original Article OBJECTIVE: To determine the optimized protocol for wet monopolar radiofrequency ablation (RFA) using a perfused-cooled electrode to induce coagulation necrosis in the ex vivo bovine liver. MATERIALS AND METHODS: Radiofrequency was applied to excised bovine livers in a monopolar mode using a 200W generator with an internally cooled electrode (groups A and B) or a perfused-cooled electrode (groups C, D, E, and F) at maximum power (150-200 W) for 10 minutes. A total of 60 ablation zones were created with six different regimens: group A - dry RFA using intra-electrode cooling; group B - dry RFA using intra-electrode cooling and a pulsing algorithm; group C - wet RFA using only interstitial hypertonic saline (HS) infusion; group D - wet RFA using interstitial HS infusion and a pulsing algorithm; group E - wet RFA using interstitial HS infusion and intra-electrode cooling; and group F - wet RFA using interstitial HS infusion, intra-electrode cooling and a pulsing algorithm. In groups C, D, E, and F, RFA was performed with the infusion of 6% HS through the perfused cooled electrode at a rate of 2 mL/minute. During RFA, we measured the tissue temperature at a distance of 15 mm from the electrode. The dimensions of the ablation zones and the changes in impedance, currents, and liver temperature during RFA were compared between these six groups. RESULTS: During RFA, the mean tissue impedances in groups A (243 ± 88 Ω) and C (252.5 ± 108 Ω) were significantly higher than those in groups B (85 ± 18.7 Ω), D (108.2 ± 85 Ω), E (70.0 ± 16.3 Ω), and F (66.5 ± 7 Ω) (p < 0.05). The mean currents in groups E and F were significantly higher than those in groups B and D, which were significantly higher than those in groups A and C (p < 0.05): 520 ± 425 mA in group A, 1163 ± 34 mA in group B, 652.5 ± 418 mA in group C, 842.5 ± 773 mA in group D, 1665 ± 295 mA in group E, and 1830 ± 109 mA in group F. The mean volumes of the ablation regions in groups E and F were significantly larger than those in the other groups (p < 0.05): 17.7 ± 5.6 cm(3) in group A, 34.5 ± 3.0 cm(3) in group B, 20.2 ± 15.6 cm(3) in group C, 36.1 ± 19.5 cm(3) in group D, 68.1 ± 12.4 cm(3) in group E, and 79.5 ± 31 cm(3) in group F. The final tissue temperatures at a distance of 15 mm from the electrode were higher in groups E and F than those in groups A, C, and D (p < 0.05): 50 ± 7.5℃ in group A, 66 ± 13.6℃ in group B, 60 ± 13.4℃ in group C, 61 ±12.7℃ in group D, 78 ± 14.2℃ in group E, and 79 ± 12.0℃ in group F. CONCLUSION: Wet monopolar RFA, using intra-electrode cooling and interstitial saline infusion, showed better performance in creating a large ablation zone than either dry RFA or wet RFA without intra-electrode cooling. The Korean Radiological Society 2004 2004-12-31 /pmc/articles/PMC2698169/ /pubmed/15637475 http://dx.doi.org/10.3348/kjr.2004.5.4.250 Text en Copyright © 2004 The Korean Radiological Society http://creativecommons.org/licenses/by-nc/3.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Lee, Jeong Min Han, Joon Koo Kim, Se Hyung Lee, Jae Young Shin, Kyung Sook Han, Chang Jin Lee, Min Woo Choi, Jun Il Choi, Byung Ihn Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers |
title | Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers |
title_full | Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers |
title_fullStr | Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers |
title_full_unstemmed | Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers |
title_short | Optimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers |
title_sort | optimization of wet radiofrequency ablation using a perfused-cooled electrode: a comparative study in ex vivo bovine livers |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2698169/ https://www.ncbi.nlm.nih.gov/pubmed/15637475 http://dx.doi.org/10.3348/kjr.2004.5.4.250 |
work_keys_str_mv | AT leejeongmin optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT hanjoonkoo optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT kimsehyung optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT leejaeyoung optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT shinkyungsook optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT hanchangjin optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT leeminwoo optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT choijunil optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers AT choibyungihn optimizationofwetradiofrequencyablationusingaperfusedcooledelectrodeacomparativestudyinexvivobovinelivers |