Inability of a Confocal Scanning Laser Doppler Flowmeter to Measure Choroidal Blood Flow in the Pig Eye

The aim of the study was to determine whether the Heidelberg Retinal Flowmeter (HRF), a confocal scanning laser Doppler flowmeter, can measure choroidal blood flow in pig eyes. An HRF was used to obtain flow maps from in vitro pig eyes under a range of perfusion flow rates (0 – 500 µL/min) under con...

Descripción completa

Detalles Bibliográficos
Autores principales: Pandav, S, Morgan, W.H, Townsend, R, Cringle, S.J, Yu, D.Y
Formato: Texto
Lenguaje:English
Publicado: Bentham Open 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2699845/
https://www.ncbi.nlm.nih.gov/pubmed/19547666
http://dx.doi.org/10.2174/1874364100802010146
Descripción
Sumario:The aim of the study was to determine whether the Heidelberg Retinal Flowmeter (HRF), a confocal scanning laser Doppler flowmeter, can measure choroidal blood flow in pig eyes. An HRF was used to obtain flow maps from in vitro pig eyes under a range of perfusion flow rates (0 – 500 µL/min) under conditions in which only the choroid was perfused. In some cases choroidal blood flow was also measured simultaneously using a conventional fiberoptic based Laser Doppler Perfusion Monitor (LDPM) which used the same laser wavelength (780 nm). The relationship between perfusion flow, HRF measured flow and LDPM measured flow was determined. HRF flow maps were also obtained in vivo as a function of focal plane setting through the retina and choroid. Across the range of perfusion flow rates through the isolated eyes there was a poor correlation with HRF measured choroidal flow and perfusion flow. In contrast, there was a strong linear relationship between perfusion flow and LDPM measured blood flow. Both in vitro and in vivo, no choroidal vessels could be visualised in the HRF flow maps, even when the focal plane was in the choroid. The HRF is unable to measure choroidal blood flow in pig eyes. This is not due to an inability of the 780 nm laser to penetrate into the choroid or due to red blood cell velocities in the choroid being higher than the measurement range of the instrument.