Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?

OBJECTIVE: An endotracheal tube (ETT) imposes work of breathing on mechanically ventilated patients. Using a bellows-in-a-box model lung, we compared the tube compensation (TC) performances of the Nellcor Puritan-Bennett 840 ventilator and of the Dräger Evita 4 ventilator. MEASUREMENTS AND RESULTS:...

Descripción completa

Detalles Bibliográficos
Autores principales: Maeda, Yoshiko, Fujino, Yuji, Uchiyama, Akinori, Taenaka, Nobuyuki, Mashimo, Takashi, Nishimura, Masaji
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC270715/
https://www.ncbi.nlm.nih.gov/pubmed/12974975
http://dx.doi.org/10.1186/cc2343
_version_ 1782121046498869248
author Maeda, Yoshiko
Fujino, Yuji
Uchiyama, Akinori
Taenaka, Nobuyuki
Mashimo, Takashi
Nishimura, Masaji
author_facet Maeda, Yoshiko
Fujino, Yuji
Uchiyama, Akinori
Taenaka, Nobuyuki
Mashimo, Takashi
Nishimura, Masaji
author_sort Maeda, Yoshiko
collection PubMed
description OBJECTIVE: An endotracheal tube (ETT) imposes work of breathing on mechanically ventilated patients. Using a bellows-in-a-box model lung, we compared the tube compensation (TC) performances of the Nellcor Puritan-Bennett 840 ventilator and of the Dräger Evita 4 ventilator. MEASUREMENTS AND RESULTS: Each ventilator was connected to the model lung. The respiratory rate of the model lung was set at 10 breaths/min with 1 s inspiratory time. Inspiratory flows were 30 or 60 l/min. A full-length 8 mm bore ETT was inserted between the ventilator circuit and the model lung. The TC was set at 0%, 10%, 50%, and 100% for both ventilators. Pressure was monitored at the airway, the trachea, and the pleura, and the data were recorded on a computer for later analysis of the delay time, of the inspiratory trigger pressure, and of the pressure–time product (PTP). The delay time was calculated as the time between the start of inspiration and minimum airway pressure, and the inspiratory trigger pressure was defined as the most negative pressure level. The same measurements were performed under pressure support ventilation of 4 and 8 cmH(2)O. The PTP increased according to the magnitude of inspiratory flow. Even with 100% TC, neither ventilator could completely compensate for the PTP imposed by the ETT. At 0% TC the PTP tended to be less with the Nellcor Puritan-Bennett 840 ventilator, while at 100% TC the PTP tended to be less with the Dräger Evita 4 ventilator. A small amount of pressure support can be equally effective to reduce the inspiratory effort compared with the TC. CONCLUSION: Although both ventilators provided effective TC, even when set to 100% TC they could not entirely compensate for a ventilator and ETT-imposed work of breathing. The effect of TC is less than that of pressure support ventilation. Physicians should be aware of this when using TC in weaning trials.
format Text
id pubmed-270715
institution National Center for Biotechnology Information
language English
publishDate 2003
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-2707152003-11-21 Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief? Maeda, Yoshiko Fujino, Yuji Uchiyama, Akinori Taenaka, Nobuyuki Mashimo, Takashi Nishimura, Masaji Crit Care Research OBJECTIVE: An endotracheal tube (ETT) imposes work of breathing on mechanically ventilated patients. Using a bellows-in-a-box model lung, we compared the tube compensation (TC) performances of the Nellcor Puritan-Bennett 840 ventilator and of the Dräger Evita 4 ventilator. MEASUREMENTS AND RESULTS: Each ventilator was connected to the model lung. The respiratory rate of the model lung was set at 10 breaths/min with 1 s inspiratory time. Inspiratory flows were 30 or 60 l/min. A full-length 8 mm bore ETT was inserted between the ventilator circuit and the model lung. The TC was set at 0%, 10%, 50%, and 100% for both ventilators. Pressure was monitored at the airway, the trachea, and the pleura, and the data were recorded on a computer for later analysis of the delay time, of the inspiratory trigger pressure, and of the pressure–time product (PTP). The delay time was calculated as the time between the start of inspiration and minimum airway pressure, and the inspiratory trigger pressure was defined as the most negative pressure level. The same measurements were performed under pressure support ventilation of 4 and 8 cmH(2)O. The PTP increased according to the magnitude of inspiratory flow. Even with 100% TC, neither ventilator could completely compensate for the PTP imposed by the ETT. At 0% TC the PTP tended to be less with the Nellcor Puritan-Bennett 840 ventilator, while at 100% TC the PTP tended to be less with the Dräger Evita 4 ventilator. A small amount of pressure support can be equally effective to reduce the inspiratory effort compared with the TC. CONCLUSION: Although both ventilators provided effective TC, even when set to 100% TC they could not entirely compensate for a ventilator and ETT-imposed work of breathing. The effect of TC is less than that of pressure support ventilation. Physicians should be aware of this when using TC in weaning trials. BioMed Central 2003 2003-08-14 /pmc/articles/PMC270715/ /pubmed/12974975 http://dx.doi.org/10.1186/cc2343 Text en Copyright © 2003 Maeda et al., licensee BioMed Central Ltd. This is an Open Access article: verbatim copying and redistribution of this article are permitted in all media for any purpose, provided this notice is preserved along with the article's original URL.
spellingShingle Research
Maeda, Yoshiko
Fujino, Yuji
Uchiyama, Akinori
Taenaka, Nobuyuki
Mashimo, Takashi
Nishimura, Masaji
Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
title Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
title_full Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
title_fullStr Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
title_full_unstemmed Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
title_short Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
title_sort does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC270715/
https://www.ncbi.nlm.nih.gov/pubmed/12974975
http://dx.doi.org/10.1186/cc2343
work_keys_str_mv AT maedayoshiko doesthetubecompensationfunctionoftwomodernmechanicalventilatorsprovideeffectiveworkofbreathingrelief
AT fujinoyuji doesthetubecompensationfunctionoftwomodernmechanicalventilatorsprovideeffectiveworkofbreathingrelief
AT uchiyamaakinori doesthetubecompensationfunctionoftwomodernmechanicalventilatorsprovideeffectiveworkofbreathingrelief
AT taenakanobuyuki doesthetubecompensationfunctionoftwomodernmechanicalventilatorsprovideeffectiveworkofbreathingrelief
AT mashimotakashi doesthetubecompensationfunctionoftwomodernmechanicalventilatorsprovideeffectiveworkofbreathingrelief
AT nishimuramasaji doesthetubecompensationfunctionoftwomodernmechanicalventilatorsprovideeffectiveworkofbreathingrelief