A retrospective randomized study to compare the energy delivered using CDE with different techniques and OZil(®) settings by different surgeons in phacoemulsification

Cumulative dissipated energy (CDE) was used with Infiniti(®) Vision System (Alcon Labs) as an energy delivery guide to compare four different phaco techniques and phaco settings. The supracapsular phaco technique and burst mode is known for efficiency and surgery is faster compared with the old phac...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Ming, Sweeney, Henry W, Luke, Becky, Chen, Mindy, Brown, Mathew
Formato: Texto
Lenguaje:English
Publicado: Dove Medical Press 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2720679/
https://www.ncbi.nlm.nih.gov/pubmed/19688027
Descripción
Sumario:Cumulative dissipated energy (CDE) was used with Infiniti(®) Vision System (Alcon Labs) as an energy delivery guide to compare four different phaco techniques and phaco settings. The supracapsular phaco technique and burst mode is known for efficiency and surgery is faster compared with the old phaco unit. In this study, we found that supracapsular phaco with burst mode had the least CDE in both cataract and nuclear sclerosis cataract with the new Infiniti(®) unit. We suggest that CDE can be used as one of the references to modify technique and setting to improve outcome for surgeons, especially for new surgeons.