Simultaneous integrated boost radiotherapy for bilateral breast: a treatment planning and dosimetric comparison for volumetric modulated arc and fixed field intensity modulated therapy
PURPOSE: A study was performed comparing dosimetric characteristics of volumetric modulated arcs (RapidArc, RA) and fixed field intensity modulated therapy (IMRT) on patients with bilateral breast carcinoma. MATERIALS AND METHODS: Plans for IMRT and RA, were optimised for 10 patients prescribing 50...
Autores principales: | , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2722660/ https://www.ncbi.nlm.nih.gov/pubmed/19630947 http://dx.doi.org/10.1186/1748-717X-4-27 |
Sumario: | PURPOSE: A study was performed comparing dosimetric characteristics of volumetric modulated arcs (RapidArc, RA) and fixed field intensity modulated therapy (IMRT) on patients with bilateral breast carcinoma. MATERIALS AND METHODS: Plans for IMRT and RA, were optimised for 10 patients prescribing 50 Gy to the breast (PTVII, 2.0 Gy/fraction) and 60 Gy to the tumour bed (PTVI, 2.4 Gy/fraction). Objectives were: for PTVs V(90%)>95%, D(max)<107%; Mean lung dose MLD<15 Gy, V(20 Gy)<22%; heart involvement was to be minimised. The MU and delivery time measured treatment efficiency. Pre-treatment dosimetry was performed using EPID and a 2D-array based methods. RESULTS: For PTVII minus PTVI, V(90% )was 97.8 ± 3.4% for RA and 94.0 ± 3.5% for IMRT (findings are reported as mean ± 1 standard deviation); D(5%)-D(95% )(homogeneity) was 7.3 ± 1.4 Gy (RA) and 11.0 ± 1.1 Gy (IMRT). Conformity index (V(95%)/V(PTVII)) was 1.10 ± 0.06 (RA) and 1.14 ± 0.09 (IMRT). MLD was <9.5 Gy for all cases on each lung, V(20 Gy )was 9.7 ± 1.3% (RA) and 12.8 ± 2.5% (IMRT) on left lung, similar for right lung. Mean dose to heart was 6.0 ± 2.7 Gy (RA) and 7.4 ± 2.5 Gy (IMRT). MU resulted in 796 ± 121 (RA) and 1398 ± 301 (IMRT); the average measured treatment time was 3.0 ± 0.1 minutes (RA) and 11.5 ± 2.0 (IMRT). From pre-treatment dosimetry, % of field area with γ <1 resulted 98.8 ± 1.3% and 99.1 ± 1.5% for RA and IMRT respectively with EPID and 99.1 ± 1.8% and 99.5 ± 1.3% with 2D-array (ΔD = 3% and DTA = 3 mm). CONCLUSION: RapidArc showed dosimetric improvements with respect to IMRT, delivery parameters confirmed its logistical advantages, pre-treatment dosimetry proved its reliability. |
---|