A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA)
Dual Energy X-ray Absorptiometry (DXA) is effective in measuring bone mineral density (BMD) in mice for early detection of osteoporosis. However, scanners designed for use with small animals (i.e. PIXImus) are very expensive. Used human DXA machines are cheaper to obtain, but analysis of scans from...
Autores principales: | , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Informa Healthcare
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2739646/ https://www.ncbi.nlm.nih.gov/pubmed/19750018 http://dx.doi.org/10.1080/15376510802499030 |
_version_ | 1782171613946445824 |
---|---|
author | Katikaneni, Ranjitha Ponnapakkam, Adharsh Miller, Eric Ponnapakkam, Tulasi Gensure, Robert C. |
author_facet | Katikaneni, Ranjitha Ponnapakkam, Adharsh Miller, Eric Ponnapakkam, Tulasi Gensure, Robert C. |
author_sort | Katikaneni, Ranjitha |
collection | PubMed |
description | Dual Energy X-ray Absorptiometry (DXA) is effective in measuring bone mineral density (BMD) in mice for early detection of osteoporosis. However, scanners designed for use with small animals (i.e. PIXImus) are very expensive. Used human DXA machines are cheaper to obtain, but analysis of scans from these instruments is operator-dependent. Obtaining reliable data depends on having a single operator analyze the scans in a blinded fashion. Scan quality is improved by excising the bone prior to scanning, which does not allow serial measurements. This study describes a novel method of analyzing lumbar spine BMD in mice using whole body DXA. This non-invasive technique has a high degree of precision and reproducibility, with good correlation between multiple observers. Inter-observer variability (0.063 ± 0.00317 g/cm(2) [mean ± SD], 5.05 [% coefficient of variation (CV)], repeat scan variability (0.063 ± 0.00364 g/cm(2) [mean ± SD], 5.94 [%CV]) were very low compared to variability between different animals (0.063 ± 0.00588 g/cm(2) [mean ± SD], 9.64 [%CV]) and variability seen in same animal over time (0.011 ± 0.00885 g/cm(2) [mean ± SD], 80.68 [%CV]). The measurement error is thus smaller than the biological variation. Accuracy was determined by comparing average peak BMD from two scans per mouse in-vivo (0.066 g/cm(2)) versus excised spine (0.065 g/cm(2)). Furthermore, correlation between bone ash weights and whole body lumbar spine BMD measurements (p < 0.0001) was highly significant. This technique thus shows a high degree of precision and accuracy, even with multiple observers, for measuring BMD in mice using a DXA machine designed for clinical use. |
format | Text |
id | pubmed-2739646 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2009 |
publisher | Informa Healthcare |
record_format | MEDLINE/PubMed |
spelling | pubmed-27396462009-09-10 A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA) Katikaneni, Ranjitha Ponnapakkam, Adharsh Miller, Eric Ponnapakkam, Tulasi Gensure, Robert C. Toxicol Mech Methods Research Article Dual Energy X-ray Absorptiometry (DXA) is effective in measuring bone mineral density (BMD) in mice for early detection of osteoporosis. However, scanners designed for use with small animals (i.e. PIXImus) are very expensive. Used human DXA machines are cheaper to obtain, but analysis of scans from these instruments is operator-dependent. Obtaining reliable data depends on having a single operator analyze the scans in a blinded fashion. Scan quality is improved by excising the bone prior to scanning, which does not allow serial measurements. This study describes a novel method of analyzing lumbar spine BMD in mice using whole body DXA. This non-invasive technique has a high degree of precision and reproducibility, with good correlation between multiple observers. Inter-observer variability (0.063 ± 0.00317 g/cm(2) [mean ± SD], 5.05 [% coefficient of variation (CV)], repeat scan variability (0.063 ± 0.00364 g/cm(2) [mean ± SD], 5.94 [%CV]) were very low compared to variability between different animals (0.063 ± 0.00588 g/cm(2) [mean ± SD], 9.64 [%CV]) and variability seen in same animal over time (0.011 ± 0.00885 g/cm(2) [mean ± SD], 80.68 [%CV]). The measurement error is thus smaller than the biological variation. Accuracy was determined by comparing average peak BMD from two scans per mouse in-vivo (0.066 g/cm(2)) versus excised spine (0.065 g/cm(2)). Furthermore, correlation between bone ash weights and whole body lumbar spine BMD measurements (p < 0.0001) was highly significant. This technique thus shows a high degree of precision and accuracy, even with multiple observers, for measuring BMD in mice using a DXA machine designed for clinical use. Informa Healthcare 2009-06-30 2009-03 /pmc/articles/PMC2739646/ /pubmed/19750018 http://dx.doi.org/10.1080/15376510802499030 Text en © 2009 Informa UK Ltd http://creativecommons.org/licenses/by/2.0/ This is an open access article distributed under the Supplemental Terms and Conditions for iOpenAccess articles published in Informa Healthcare journals (http://www.informaworld.com/mpp/uploads/iopenaccess_tcs.pdf) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Katikaneni, Ranjitha Ponnapakkam, Adharsh Miller, Eric Ponnapakkam, Tulasi Gensure, Robert C. A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA) |
title | A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA) |
title_full | A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA) |
title_fullStr | A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA) |
title_full_unstemmed | A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA) |
title_short | A new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy X-ray absorptiometry (DXA) |
title_sort | new technique for precisely and accurately measuring lumbar spine bone mineral density in mice using clinical dual energy x-ray absorptiometry (dxa) |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2739646/ https://www.ncbi.nlm.nih.gov/pubmed/19750018 http://dx.doi.org/10.1080/15376510802499030 |
work_keys_str_mv | AT katikaneniranjitha anewtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT ponnapakkamadharsh anewtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT millereric anewtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT ponnapakkamtulasi anewtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT gensurerobertc anewtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT katikaneniranjitha newtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT ponnapakkamadharsh newtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT millereric newtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT ponnapakkamtulasi newtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa AT gensurerobertc newtechniqueforpreciselyandaccuratelymeasuringlumbarspinebonemineraldensityinmiceusingclinicaldualenergyxrayabsorptiometrydxa |