Cargando…

Aerosol delivery by an ultrasonic nebulizer during different mechanical ventilation settings in a lung model – a pilot study

Aerosol delivery of Iloprost is a promising therapeutic approach. The aim of this study was to determine the output of an ultrasonic nebulizer in different ventilation set-ups at the tip of different endotracheal tubes. METHOD: In set-up A, an ultrasonic nebulizer was connected directly to the endot...

Descripción completa

Detalles Bibliográficos
Autores principales: Winterhalter, Michael, Bund, Michael, Khaladj, Nawid, Hagl, Christian, Simon, Andre, Hoy, Ludwig, Piepenbrock, Siegfried, Rahe-Meyer, Niels
Formato: Texto
Lenguaje:English
Publicado: Dove Medical Press 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2761175/
https://www.ncbi.nlm.nih.gov/pubmed/19920888
Descripción
Sumario:Aerosol delivery of Iloprost is a promising therapeutic approach. The aim of this study was to determine the output of an ultrasonic nebulizer in different ventilation set-ups at the tip of different endotracheal tubes. METHOD: In set-up A, an ultrasonic nebulizer was connected directly to the endotracheal tube. In set-up B, the nebulizer was incorporated into the inspiratory limb of the ventilator circuit; a bypass arrangement allowed to selectively direct the exspiratory air discharged from the model lung. The test lungs were ventilated through a standard endotracheal tube (ET) and through a double-lumen tube (DLT). The nebulizer was filled with 5 ml of a Tc-99m 0.9%-NaCl solution. After nebulization, distribution of radioactivity was detected by gamma scintigraphy. RESULTS: Set-up A, ventilation in volume-controlled mode (VCV) via an ET: Delivered dose (1.61 ± 0.41 ml), nebulization time 10.13 ± 1.71 min. Set-up A, pressure-controlled ventilation (PCV), via a DLT: Delivered dose (1.33 ± 0.88 ml), nebulization time 13.27 ± 2.58 min. Set-up B, VCV mode via an ET: Delivered dose (1.57 ± 0.44 ml), nebulization time (25.9 ± 3.8 min). Set-up B, PCV mode, via a DLT: Delivered dose (1.3 ± 0.17 ml), nebulization time (25.6 ± 4.0 min). Set-up B did not yield a significantly higher output (p < 0.05), but the nebulization time was significantly longer (p > 0.05) compared with set-up A. CONCLUSION: Set-ups which involve connecting the nebulizer directly to an ET or a DLT exhibit sufficient output of aerosol and short nebulization times.