Cargando…
Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator
An Image-Guided Radiotherapy–capable linear accelerator (Elekta Synergy) was installed at our hospital, which is equipped with a kV x-ray volume imaging (XVI) system and electronic portal imaging device (iViewGT). The objective of this presentation is to describe the results of commissioning measure...
Autores principales: | , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2772033/ https://www.ncbi.nlm.nih.gov/pubmed/19893694 http://dx.doi.org/10.4103/0971-6203.41276 |
_version_ | 1782173797730746368 |
---|---|
author | Muralidhar, K. R. Murthy, P. Narayana Kumar, Rajneesh |
author_facet | Muralidhar, K. R. Murthy, P. Narayana Kumar, Rajneesh |
author_sort | Muralidhar, K. R. |
collection | PubMed |
description | An Image-Guided Radiotherapy–capable linear accelerator (Elekta Synergy) was installed at our hospital, which is equipped with a kV x-ray volume imaging (XVI) system and electronic portal imaging device (iViewGT). The objective of this presentation is to describe the results of commissioning measurements carried out on the XVI facility to verify the manufacturer's specifications and also to evolve a QA schedule which can be used to test its performance routinely. The QA program consists of a series of tests (safety features, geometric accuracy, and image quality). These tests were found to be useful to assess the performance of the XVI system and also proved that XVI system is very suitable for image-guided high-precision radiation therapy. |
format | Text |
id | pubmed-2772033 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | Medknow Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-27720332009-11-05 Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator Muralidhar, K. R. Murthy, P. Narayana Kumar, Rajneesh J Med Phys Technical Note An Image-Guided Radiotherapy–capable linear accelerator (Elekta Synergy) was installed at our hospital, which is equipped with a kV x-ray volume imaging (XVI) system and electronic portal imaging device (iViewGT). The objective of this presentation is to describe the results of commissioning measurements carried out on the XVI facility to verify the manufacturer's specifications and also to evolve a QA schedule which can be used to test its performance routinely. The QA program consists of a series of tests (safety features, geometric accuracy, and image quality). These tests were found to be useful to assess the performance of the XVI system and also proved that XVI system is very suitable for image-guided high-precision radiation therapy. Medknow Publications 2008 /pmc/articles/PMC2772033/ /pubmed/19893694 http://dx.doi.org/10.4103/0971-6203.41276 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Note Muralidhar, K. R. Murthy, P. Narayana Kumar, Rajneesh Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator |
title | Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator |
title_full | Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator |
title_fullStr | Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator |
title_full_unstemmed | Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator |
title_short | Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator |
title_sort | commissioning and quality assurance of the x-ray volume imaging system of an image-guided radiotherapy capable linear accelerator |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2772033/ https://www.ncbi.nlm.nih.gov/pubmed/19893694 http://dx.doi.org/10.4103/0971-6203.41276 |
work_keys_str_mv | AT muralidharkr commissioningandqualityassuranceofthexrayvolumeimagingsystemofanimageguidedradiotherapycapablelinearaccelerator AT murthypnarayana commissioningandqualityassuranceofthexrayvolumeimagingsystemofanimageguidedradiotherapycapablelinearaccelerator AT kumarrajneesh commissioningandqualityassuranceofthexrayvolumeimagingsystemofanimageguidedradiotherapycapablelinearaccelerator |