Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds

INTRODUCTION: Tissue oxygen saturation (StO(2)) and the vascular occlusion test (VOT) can identify tissue hypoperfusion in trauma and sepsis. However, the technique is neither standardized nor uses the same monitoring site. We hypothesized that baseline and VOT StO(2 )would be different in the forea...

Descripción completa

Detalles Bibliográficos
Autores principales: Gómez, Hernando, Mesquida, Jaume, Simon, Peter, Kim, Hyung Kook, Puyana, Juan C, Ince, Can, Pinsky, Michael R
Formato: Texto
Lenguaje:English
Publicado: BioMed Central 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2786105/
https://www.ncbi.nlm.nih.gov/pubmed/19951387
http://dx.doi.org/10.1186/cc8001
_version_ 1782174860161581056
author Gómez, Hernando
Mesquida, Jaume
Simon, Peter
Kim, Hyung Kook
Puyana, Juan C
Ince, Can
Pinsky, Michael R
author_facet Gómez, Hernando
Mesquida, Jaume
Simon, Peter
Kim, Hyung Kook
Puyana, Juan C
Ince, Can
Pinsky, Michael R
author_sort Gómez, Hernando
collection PubMed
description INTRODUCTION: Tissue oxygen saturation (StO(2)) and the vascular occlusion test (VOT) can identify tissue hypoperfusion in trauma and sepsis. However, the technique is neither standardized nor uses the same monitoring site. We hypothesized that baseline and VOT StO(2 )would be different in the forearm (F) and thenar eminence (TH) and that different minimal StO(2 )values during the VOT would result in different reoxygenation rates (ReO(2)). METHODS: StO(2 )and its change during the VOT were simultaneously measured in the F and TH, with 15 mm and 25 mm probes, using the 325 InSpectra monitor in 18 healthy, adult volunteers. Two VOTs were done to a threshold thenar StO(2 )of 40% interchanging the 15 mm and 25 mm probes between sites. Two additional VOTs were done to thresholds of 50% and 30%. Baseline StO(2 )(BaseO(2)), the deoxygenation rate (DeO(2)) and ReO(2 )were compared between sites, probes and (%O(2)/minute) thresholds. Results are presented as the median (interquartile range), P-value. RESULTS: BaseO(2), DeO(2), ReO(2), area under the curve and hyperemia duration values were different when comparing TH vs. F and 15 mm vs. 25 mm probes. ReO(2 )was different between different thresholds for the TH and 15 mm probes. TH(15 mm )vs. F(15 mm): BaseO(2), 90.4 (85.2, 93.5) vs. 85.2 (80.7, 90.2), P = 0.031; DO(2), -12.1 (-16.2, -11.3) vs. -8.5 (-10.3, -7.8), P = 0.011; ReO(2), 297.2 (213.7, 328.6), P < 0.0001; 15 mm vs. 25 mm probe: BaseO(2), 97.2 (89.4, 94.7) vs. 87.3 (81.7, 90.9), P = 0.016; DeO(2), -18.0 (-24.1, -14.8) vs. -9.9 (-15.3, -6.5), P < 0.0001; and ReO(2), 401.6 (331.7, 543.2) vs. 160.5 (132.3, 366.9), P = 0.012, respectively. TH(15 mm )vs. TH(25 mm): BaseO(2), P = 0.020; DeO(2), P < 0.0001; and ReO(2), P < 0.0001. Threshold StO(2 )values (15 mm probe only): ReO(2), P = 0.003; DeO(2), P = 0.60. ReO(2 )at 40% and 50% StO(2 )thresholds, P = 0.01. CONCLUSIONS: BaseO(2), DeO(2 )and ReO(2 )were different when measured in different anatomical sites (F and TH) and with different probe sizes, and ReO(2 )was different with differing VOT release StO(2 )threshold values. Thus, standardization of the site, probe and VOT challenge need to be stipulated when reporting data.
format Text
id pubmed-2786105
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-27861052009-12-02 Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds Gómez, Hernando Mesquida, Jaume Simon, Peter Kim, Hyung Kook Puyana, Juan C Ince, Can Pinsky, Michael R Crit Care Research INTRODUCTION: Tissue oxygen saturation (StO(2)) and the vascular occlusion test (VOT) can identify tissue hypoperfusion in trauma and sepsis. However, the technique is neither standardized nor uses the same monitoring site. We hypothesized that baseline and VOT StO(2 )would be different in the forearm (F) and thenar eminence (TH) and that different minimal StO(2 )values during the VOT would result in different reoxygenation rates (ReO(2)). METHODS: StO(2 )and its change during the VOT were simultaneously measured in the F and TH, with 15 mm and 25 mm probes, using the 325 InSpectra monitor in 18 healthy, adult volunteers. Two VOTs were done to a threshold thenar StO(2 )of 40% interchanging the 15 mm and 25 mm probes between sites. Two additional VOTs were done to thresholds of 50% and 30%. Baseline StO(2 )(BaseO(2)), the deoxygenation rate (DeO(2)) and ReO(2 )were compared between sites, probes and (%O(2)/minute) thresholds. Results are presented as the median (interquartile range), P-value. RESULTS: BaseO(2), DeO(2), ReO(2), area under the curve and hyperemia duration values were different when comparing TH vs. F and 15 mm vs. 25 mm probes. ReO(2 )was different between different thresholds for the TH and 15 mm probes. TH(15 mm )vs. F(15 mm): BaseO(2), 90.4 (85.2, 93.5) vs. 85.2 (80.7, 90.2), P = 0.031; DO(2), -12.1 (-16.2, -11.3) vs. -8.5 (-10.3, -7.8), P = 0.011; ReO(2), 297.2 (213.7, 328.6), P < 0.0001; 15 mm vs. 25 mm probe: BaseO(2), 97.2 (89.4, 94.7) vs. 87.3 (81.7, 90.9), P = 0.016; DeO(2), -18.0 (-24.1, -14.8) vs. -9.9 (-15.3, -6.5), P < 0.0001; and ReO(2), 401.6 (331.7, 543.2) vs. 160.5 (132.3, 366.9), P = 0.012, respectively. TH(15 mm )vs. TH(25 mm): BaseO(2), P = 0.020; DeO(2), P < 0.0001; and ReO(2), P < 0.0001. Threshold StO(2 )values (15 mm probe only): ReO(2), P = 0.003; DeO(2), P = 0.60. ReO(2 )at 40% and 50% StO(2 )thresholds, P = 0.01. CONCLUSIONS: BaseO(2), DeO(2 )and ReO(2 )were different when measured in different anatomical sites (F and TH) and with different probe sizes, and ReO(2 )was different with differing VOT release StO(2 )threshold values. Thus, standardization of the site, probe and VOT challenge need to be stipulated when reporting data. BioMed Central 2009 2009-11-30 /pmc/articles/PMC2786105/ /pubmed/19951387 http://dx.doi.org/10.1186/cc8001 Text en Copyright ©2009 BioMed Central Ltd
spellingShingle Research
Gómez, Hernando
Mesquida, Jaume
Simon, Peter
Kim, Hyung Kook
Puyana, Juan C
Ince, Can
Pinsky, Michael R
Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds
title Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds
title_full Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds
title_fullStr Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds
title_full_unstemmed Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds
title_short Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds
title_sort characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2786105/
https://www.ncbi.nlm.nih.gov/pubmed/19951387
http://dx.doi.org/10.1186/cc8001
work_keys_str_mv AT gomezhernando characterizationoftissueoxygensaturationandthevascularocclusiontestinfluenceofmeasurementsitesprobesizesanddeflationthresholds
AT mesquidajaume characterizationoftissueoxygensaturationandthevascularocclusiontestinfluenceofmeasurementsitesprobesizesanddeflationthresholds
AT simonpeter characterizationoftissueoxygensaturationandthevascularocclusiontestinfluenceofmeasurementsitesprobesizesanddeflationthresholds
AT kimhyungkook characterizationoftissueoxygensaturationandthevascularocclusiontestinfluenceofmeasurementsitesprobesizesanddeflationthresholds
AT puyanajuanc characterizationoftissueoxygensaturationandthevascularocclusiontestinfluenceofmeasurementsitesprobesizesanddeflationthresholds
AT incecan characterizationoftissueoxygensaturationandthevascularocclusiontestinfluenceofmeasurementsitesprobesizesanddeflationthresholds
AT pinskymichaelr characterizationoftissueoxygensaturationandthevascularocclusiontestinfluenceofmeasurementsitesprobesizesanddeflationthresholds