Cargando…
Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India
A Helical Tomotherapy™ (HT) Hi-Art II (TomoTherapy, Inc., Madison, WI, USA) has been one of the important innovations to help deliver IMRT with image guidance. On-board, mega voltage computed tomography (MVCT) detectors are used for imaging and dosimetric purpose. The two objectives of this study ar...
Autores principales: | , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2805893/ https://www.ncbi.nlm.nih.gov/pubmed/20098540 http://dx.doi.org/10.4103/0971-6203.51933 |
_version_ | 1782176240020488192 |
---|---|
author | Kinhikar, Rajesh A. Master, Zubin Dhote, Dipak S. Deshpande, Deepak D. |
author_facet | Kinhikar, Rajesh A. Master, Zubin Dhote, Dipak S. Deshpande, Deepak D. |
author_sort | Kinhikar, Rajesh A. |
collection | PubMed |
description | A Helical Tomotherapy™ (HT) Hi-Art II (TomoTherapy, Inc., Madison, WI, USA) has been one of the important innovations to help deliver IMRT with image guidance. On-board, mega voltage computed tomography (MVCT) detectors are used for imaging and dosimetric purpose. The two objectives of this study are: (i) To estimate the dosimetric and general capability (TomoImage registration, reconstruction, contrast and spatial resolution, artifacts-free image and dose in TomoImage) of on-board MVCT detectors. (ii) To measure the dosimetric parameters (output and energy) following major repair. The MVCT detectors also estimated the rotational output constancy well. During this study, dosimetric tests were repeated after replacing MVCT detectors and the target. fixed-gantry/fixed-couch measurements were measured daily to investigate; the system stability. Thermoluminescense dosimeter (TLD) was used during both the measurements subsequently. The MVCT image quality with old and new detectors was comparable and hence acceptable clinically. The spatial resolution was optimal and the dose during TomoImage was 2 cGy (well within the manufacturer tolerance of 4 cGy). The results of lateral beam profiles showed an excellent agreement between the two normalized plots. The output from the rotational procedure revealed 99.7% while the energy was consistent over a period of twelve months. The Hi-Art II system has maintained its calibration to within +/− 2% and energy to within +/− 1.5% over the initial twelve-month period. Based on the periodic measurements for rotational output and consistency in the lateral beam profile shape, the on-board detector proved to be a viable dosimetric quality assurance tool for IMRT with Tomotherapy. Tomotherapy was stable from the dosimetric point of view during the twelve-month period. |
format | Text |
id | pubmed-2805893 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2009 |
publisher | Medknow Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-28058932010-01-22 Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India Kinhikar, Rajesh A. Master, Zubin Dhote, Dipak S. Deshpande, Deepak D. J Med Phys Original Article A Helical Tomotherapy™ (HT) Hi-Art II (TomoTherapy, Inc., Madison, WI, USA) has been one of the important innovations to help deliver IMRT with image guidance. On-board, mega voltage computed tomography (MVCT) detectors are used for imaging and dosimetric purpose. The two objectives of this study are: (i) To estimate the dosimetric and general capability (TomoImage registration, reconstruction, contrast and spatial resolution, artifacts-free image and dose in TomoImage) of on-board MVCT detectors. (ii) To measure the dosimetric parameters (output and energy) following major repair. The MVCT detectors also estimated the rotational output constancy well. During this study, dosimetric tests were repeated after replacing MVCT detectors and the target. fixed-gantry/fixed-couch measurements were measured daily to investigate; the system stability. Thermoluminescense dosimeter (TLD) was used during both the measurements subsequently. The MVCT image quality with old and new detectors was comparable and hence acceptable clinically. The spatial resolution was optimal and the dose during TomoImage was 2 cGy (well within the manufacturer tolerance of 4 cGy). The results of lateral beam profiles showed an excellent agreement between the two normalized plots. The output from the rotational procedure revealed 99.7% while the energy was consistent over a period of twelve months. The Hi-Art II system has maintained its calibration to within +/− 2% and energy to within +/− 1.5% over the initial twelve-month period. Based on the periodic measurements for rotational output and consistency in the lateral beam profile shape, the on-board detector proved to be a viable dosimetric quality assurance tool for IMRT with Tomotherapy. Tomotherapy was stable from the dosimetric point of view during the twelve-month period. Medknow Publications 2009 /pmc/articles/PMC2805893/ /pubmed/20098540 http://dx.doi.org/10.4103/0971-6203.51933 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Kinhikar, Rajesh A. Master, Zubin Dhote, Dipak S. Deshpande, Deepak D. Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India |
title | Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India |
title_full | Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India |
title_fullStr | Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India |
title_full_unstemmed | Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India |
title_short | Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first Helical Hi-Art II tomotherapy machine in India |
title_sort | initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical hi-art ii tomotherapy machine in india |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2805893/ https://www.ncbi.nlm.nih.gov/pubmed/20098540 http://dx.doi.org/10.4103/0971-6203.51933 |
work_keys_str_mv | AT kinhikarrajesha initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia AT masterzubin initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia AT dhotedipaks initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia AT deshpandedeepakd initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia |