Cargando…

Commissioning experience and quality assurance of helical tomotherapy machines

A helical tomotherapy machine combines a straight 6 MV linear accelerator mounted on a ring gantry with CT technology for image-guided intensity-modulated radiation therapy (IMRT) treatment. A fan beam created by the collimator and jaws produces a maximum of 40 × 5 cm(2) field size at the isocenter....

Descripción completa

Detalles Bibliográficos
Autores principales: Sen, Amarjit, West, Matthew K.
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807140/
https://www.ncbi.nlm.nih.gov/pubmed/20098548
http://dx.doi.org/10.4103/0971-6203.56078
_version_ 1782176382585929728
author Sen, Amarjit
West, Matthew K.
author_facet Sen, Amarjit
West, Matthew K.
author_sort Sen, Amarjit
collection PubMed
description A helical tomotherapy machine combines a straight 6 MV linear accelerator mounted on a ring gantry with CT technology for image-guided intensity-modulated radiation therapy (IMRT) treatment. A fan beam created by the collimator and jaws produces a maximum of 40 × 5 cm(2) field size at the isocenter. The gantry and hence the fan beam rotates at a constant speed while the couch moves linearly into the gantry bore, thus producing a helical delivery. The beam is modulated by a 64-leaf binary multileaf collimator (MLC), which enables IMRT treatment. The linac can be operated at a lower voltage (3.5 MV) and dose rate to produce megavoltage CT images, which are used for image-guided patient setup. We have installed two such units since 2004 and treated more than 2000 patients. The machine comes “precommissioned” from the manufacturer, and the beam characteristics and IMRT plans on phantom are measured and compared with manufacturer's data after acceptance tests are performed on site. Our experience with commissioning the machines and periodic quality assurance with tolerance limits for optimal performance are described.
format Text
id pubmed-2807140
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-28071402010-01-22 Commissioning experience and quality assurance of helical tomotherapy machines Sen, Amarjit West, Matthew K. J Med Phys Invited Paper A helical tomotherapy machine combines a straight 6 MV linear accelerator mounted on a ring gantry with CT technology for image-guided intensity-modulated radiation therapy (IMRT) treatment. A fan beam created by the collimator and jaws produces a maximum of 40 × 5 cm(2) field size at the isocenter. The gantry and hence the fan beam rotates at a constant speed while the couch moves linearly into the gantry bore, thus producing a helical delivery. The beam is modulated by a 64-leaf binary multileaf collimator (MLC), which enables IMRT treatment. The linac can be operated at a lower voltage (3.5 MV) and dose rate to produce megavoltage CT images, which are used for image-guided patient setup. We have installed two such units since 2004 and treated more than 2000 patients. The machine comes “precommissioned” from the manufacturer, and the beam characteristics and IMRT plans on phantom are measured and compared with manufacturer's data after acceptance tests are performed on site. Our experience with commissioning the machines and periodic quality assurance with tolerance limits for optimal performance are described. Medknow Publications 2009 /pmc/articles/PMC2807140/ /pubmed/20098548 http://dx.doi.org/10.4103/0971-6203.56078 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Invited Paper
Sen, Amarjit
West, Matthew K.
Commissioning experience and quality assurance of helical tomotherapy machines
title Commissioning experience and quality assurance of helical tomotherapy machines
title_full Commissioning experience and quality assurance of helical tomotherapy machines
title_fullStr Commissioning experience and quality assurance of helical tomotherapy machines
title_full_unstemmed Commissioning experience and quality assurance of helical tomotherapy machines
title_short Commissioning experience and quality assurance of helical tomotherapy machines
title_sort commissioning experience and quality assurance of helical tomotherapy machines
topic Invited Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807140/
https://www.ncbi.nlm.nih.gov/pubmed/20098548
http://dx.doi.org/10.4103/0971-6203.56078
work_keys_str_mv AT senamarjit commissioningexperienceandqualityassuranceofhelicaltomotherapymachines
AT westmatthewk commissioningexperienceandqualityassuranceofhelicaltomotherapymachines