Cargando…
The role of Cobalt-60 in modern radiation therapy: Dose delivery and image guidance
The advances in modern radiation therapy with techniques such as intensity-modulated radiation therapy and image-guided radiation therapy (IMRT and IGRT) have been limited almost exclusively to linear accelerators. Investigations of modern Cobalt-60 (Co-60) radiation delivery in the context of IMRT...
Autores principales: | , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807677/ https://www.ncbi.nlm.nih.gov/pubmed/20098559 http://dx.doi.org/10.4103/0971-6203.54846 |
Sumario: | The advances in modern radiation therapy with techniques such as intensity-modulated radiation therapy and image-guided radiation therapy (IMRT and IGRT) have been limited almost exclusively to linear accelerators. Investigations of modern Cobalt-60 (Co-60) radiation delivery in the context of IMRT and IGRT have been very sparse, and have been limited mainly to computer-modeling and treatment-planning exercises. In this paper, we report on the results of experiments using a tomotherapy benchtop apparatus attached to a conventional Co-60 unit. We show that conformal dose delivery is possible and also that Co-60 can be used as the radiation source in megavoltage computed tomography imaging. These results complement our modeling studies of Co-60 tomotherapy and provide a strong motivation for continuing development of modern Cobalt-60 treatment devices. |
---|