First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™)

A new model of Leksell Gamma Knife® (LGK), known as Perfexion™ (LGK PFX), was introduced by Elekta Instrument, AB, Sweden, in 2006. This model has a radically different design from the earlier models U, B, C and 4C. Dosimetric characteristics of LGK PFX, technical differences between LGK PFX and LGK...

Descripción completa

Detalles Bibliográficos
Autores principales: Bhatnagar, Jagdish P., Novotny, Josef, Niranjan, Ajay, Kondziolka1, Douglas, Flickinger, John, Lunsford, Dade, Huq, M. Saiful
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807679/
https://www.ncbi.nlm.nih.gov/pubmed/20098561
http://dx.doi.org/10.4103/0971-6203.54848
_version_ 1782176419933061120
author Bhatnagar, Jagdish P.
Novotny, Josef
Niranjan, Ajay
Kondziolka1, Douglas
Flickinger, John
Lunsford, Dade
Huq, M. Saiful
author_facet Bhatnagar, Jagdish P.
Novotny, Josef
Niranjan, Ajay
Kondziolka1, Douglas
Flickinger, John
Lunsford, Dade
Huq, M. Saiful
author_sort Bhatnagar, Jagdish P.
collection PubMed
description A new model of Leksell Gamma Knife® (LGK), known as Perfexion™ (LGK PFX), was introduced by Elekta Instrument, AB, Sweden, in 2006. This model has a radically different design from the earlier models U, B, C and 4C. Dosimetric characteristics of LGK PFX, technical differences between LGK PFX and LGK 4C, experience gained with acceptance testing and commissioning of the LGK PFX, and comparison between LGK PFX and LGK 4C are presented in this study. Excellent agreement is found between the manufacturers recommended values of absorbed dose rate, relative output factors for 4 and 8 mm collimators, coincidence of mechanical and dosimetric isocenter, FWHM for beam profiles for various collimators and those reported in the present study. Excellent agreement is also found between the dosimetric characteristics of LGK PFX and LGK 4C for the 4 and 8 mm collimators. Examples of clinical cases treated with LGK PFX and impact of LGK PFX on workflow and dosimetric conformity of treatment planning is also given. The set up and treatment of patients on the LGK PFX is much more efficient since it is a fully automated system. The system also provides more options to generate plan with high dosimetric conformity.
format Text
id pubmed-2807679
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-28076792010-01-22 First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™) Bhatnagar, Jagdish P. Novotny, Josef Niranjan, Ajay Kondziolka1, Douglas Flickinger, John Lunsford, Dade Huq, M. Saiful J Med Phys Invited Paper A new model of Leksell Gamma Knife® (LGK), known as Perfexion™ (LGK PFX), was introduced by Elekta Instrument, AB, Sweden, in 2006. This model has a radically different design from the earlier models U, B, C and 4C. Dosimetric characteristics of LGK PFX, technical differences between LGK PFX and LGK 4C, experience gained with acceptance testing and commissioning of the LGK PFX, and comparison between LGK PFX and LGK 4C are presented in this study. Excellent agreement is found between the manufacturers recommended values of absorbed dose rate, relative output factors for 4 and 8 mm collimators, coincidence of mechanical and dosimetric isocenter, FWHM for beam profiles for various collimators and those reported in the present study. Excellent agreement is also found between the dosimetric characteristics of LGK PFX and LGK 4C for the 4 and 8 mm collimators. Examples of clinical cases treated with LGK PFX and impact of LGK PFX on workflow and dosimetric conformity of treatment planning is also given. The set up and treatment of patients on the LGK PFX is much more efficient since it is a fully automated system. The system also provides more options to generate plan with high dosimetric conformity. Medknow Publications 2009 /pmc/articles/PMC2807679/ /pubmed/20098561 http://dx.doi.org/10.4103/0971-6203.54848 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Invited Paper
Bhatnagar, Jagdish P.
Novotny, Josef
Niranjan, Ajay
Kondziolka1, Douglas
Flickinger, John
Lunsford, Dade
Huq, M. Saiful
First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™)
title First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™)
title_full First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™)
title_fullStr First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™)
title_full_unstemmed First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™)
title_short First year experience with newly developed Leksell Gamma Knife(®) Perfexion(™)
title_sort first year experience with newly developed leksell gamma knife(®) perfexion(™)
topic Invited Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807679/
https://www.ncbi.nlm.nih.gov/pubmed/20098561
http://dx.doi.org/10.4103/0971-6203.54848
work_keys_str_mv AT bhatnagarjagdishp firstyearexperiencewithnewlydevelopedleksellgammaknifeperfexion
AT novotnyjosef firstyearexperiencewithnewlydevelopedleksellgammaknifeperfexion
AT niranjanajay firstyearexperiencewithnewlydevelopedleksellgammaknifeperfexion
AT kondziolka1douglas firstyearexperiencewithnewlydevelopedleksellgammaknifeperfexion
AT flickingerjohn firstyearexperiencewithnewlydevelopedleksellgammaknifeperfexion
AT lunsforddade firstyearexperiencewithnewlydevelopedleksellgammaknifeperfexion
AT huqmsaiful firstyearexperiencewithnewlydevelopedleksellgammaknifeperfexion