Verification of tomotherapy dose delivery

Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% a...

Descripción completa

Detalles Bibliográficos
Autores principales: Pelagade, S. M., Paliwal, B. R.
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807687/
https://www.ncbi.nlm.nih.gov/pubmed/20098569
http://dx.doi.org/10.4103/0971-6203.54856
Descripción
Sumario:Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% and 1.67 ± 1.5% for head and neck, prostate, pelvis-abdomen sites, and for all other sites. The spatial agreement between the calculated and the measured film dose distributions was evaluated using the gamma metric distribution. The average frequency versus gamma interval was plotted as a bar graph to quantify the gamma index variation inside the region of interest for each body site.