Verification of tomotherapy dose delivery
Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% a...
Autores principales: | , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807687/ https://www.ncbi.nlm.nih.gov/pubmed/20098569 http://dx.doi.org/10.4103/0971-6203.54856 |
_version_ | 1782176421979881472 |
---|---|
author | Pelagade, S. M. Paliwal, B. R. |
author_facet | Pelagade, S. M. Paliwal, B. R. |
author_sort | Pelagade, S. M. |
collection | PubMed |
description | Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% and 1.67 ± 1.5% for head and neck, prostate, pelvis-abdomen sites, and for all other sites. The spatial agreement between the calculated and the measured film dose distributions was evaluated using the gamma metric distribution. The average frequency versus gamma interval was plotted as a bar graph to quantify the gamma index variation inside the region of interest for each body site. |
format | Text |
id | pubmed-2807687 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2009 |
publisher | Medknow Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-28076872010-01-22 Verification of tomotherapy dose delivery Pelagade, S. M. Paliwal, B. R. J Med Phys Technical Note Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% and 1.67 ± 1.5% for head and neck, prostate, pelvis-abdomen sites, and for all other sites. The spatial agreement between the calculated and the measured film dose distributions was evaluated using the gamma metric distribution. The average frequency versus gamma interval was plotted as a bar graph to quantify the gamma index variation inside the region of interest for each body site. Medknow Publications 2009 /pmc/articles/PMC2807687/ /pubmed/20098569 http://dx.doi.org/10.4103/0971-6203.54856 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Note Pelagade, S. M. Paliwal, B. R. Verification of tomotherapy dose delivery |
title | Verification of tomotherapy dose delivery |
title_full | Verification of tomotherapy dose delivery |
title_fullStr | Verification of tomotherapy dose delivery |
title_full_unstemmed | Verification of tomotherapy dose delivery |
title_short | Verification of tomotherapy dose delivery |
title_sort | verification of tomotherapy dose delivery |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807687/ https://www.ncbi.nlm.nih.gov/pubmed/20098569 http://dx.doi.org/10.4103/0971-6203.54856 |
work_keys_str_mv | AT pelagadesm verificationoftomotherapydosedelivery AT paliwalbr verificationoftomotherapydosedelivery |