Verification of tomotherapy dose delivery

Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% a...

Descripción completa

Detalles Bibliográficos
Autores principales: Pelagade, S. M., Paliwal, B. R.
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2009
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807687/
https://www.ncbi.nlm.nih.gov/pubmed/20098569
http://dx.doi.org/10.4103/0971-6203.54856
_version_ 1782176421979881472
author Pelagade, S. M.
Paliwal, B. R.
author_facet Pelagade, S. M.
Paliwal, B. R.
author_sort Pelagade, S. M.
collection PubMed
description Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% and 1.67 ± 1.5% for head and neck, prostate, pelvis-abdomen sites, and for all other sites. The spatial agreement between the calculated and the measured film dose distributions was evaluated using the gamma metric distribution. The average frequency versus gamma interval was plotted as a bar graph to quantify the gamma index variation inside the region of interest for each body site.
format Text
id pubmed-2807687
institution National Center for Biotechnology Information
language English
publishDate 2009
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-28076872010-01-22 Verification of tomotherapy dose delivery Pelagade, S. M. Paliwal, B. R. J Med Phys Technical Note Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% and 1.67 ± 1.5% for head and neck, prostate, pelvis-abdomen sites, and for all other sites. The spatial agreement between the calculated and the measured film dose distributions was evaluated using the gamma metric distribution. The average frequency versus gamma interval was plotted as a bar graph to quantify the gamma index variation inside the region of interest for each body site. Medknow Publications 2009 /pmc/articles/PMC2807687/ /pubmed/20098569 http://dx.doi.org/10.4103/0971-6203.54856 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Note
Pelagade, S. M.
Paliwal, B. R.
Verification of tomotherapy dose delivery
title Verification of tomotherapy dose delivery
title_full Verification of tomotherapy dose delivery
title_fullStr Verification of tomotherapy dose delivery
title_full_unstemmed Verification of tomotherapy dose delivery
title_short Verification of tomotherapy dose delivery
title_sort verification of tomotherapy dose delivery
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2807687/
https://www.ncbi.nlm.nih.gov/pubmed/20098569
http://dx.doi.org/10.4103/0971-6203.54856
work_keys_str_mv AT pelagadesm verificationoftomotherapydosedelivery
AT paliwalbr verificationoftomotherapydosedelivery