Cargando…
Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device
Background. We developed a completely closed system based on gravity separation without centrifugation steps for separation of whole blood. With this new system we compared quality and stability of the processed blood components (PRC and plasma) with respect to classical preparation. Furthermore the...
Autores principales: | , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2009
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2821640/ https://www.ncbi.nlm.nih.gov/pubmed/20169121 http://dx.doi.org/10.1155/2009/175234 |
_version_ | 1782177457246306304 |
---|---|
author | Brune, T. Hannemann-Pohl, K. Nißle, K. Ecker, N. Garritsen, H. |
author_facet | Brune, T. Hannemann-Pohl, K. Nißle, K. Ecker, N. Garritsen, H. |
author_sort | Brune, T. |
collection | PubMed |
description | Background. We developed a completely closed system based on gravity separation without centrifugation steps for separation of whole blood. With this new system we compared quality and stability of the processed blood components (PRC and plasma) with respect to classical preparation. Furthermore the cost-effectiveness of this hollow fibre system was evaluated. Study Design and Methods. Whole blood collections of 15 regular blood donors were used for component preparation using the U shaped hollow fibre filter device. Results were compared to 15 whole blood preparations using centrifugation. The following parameters were evaluated: total hemoglobin, leukocyte counts, the serum concentration of total protein, lactate dehydrogenase (LDH) and potassium. Furthermore ATIII, vWF and F VIII were analyzed at different timepoints. Results. packed red cells: the data directly after separation and after 42 days of storage are in line with the guidelines of the council of Europe. Plasma. all plasma quality data are in line with the guidelines of the council of Europe for quality assurance of plasma, except for a low protein amount (factor 0.75). Conclusion. Separation of whole blood on a clinical scale in this new closed system is feasible, however the plasma protein content must be optimized. |
format | Text |
id | pubmed-2821640 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2009 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-28216402010-02-18 Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device Brune, T. Hannemann-Pohl, K. Nißle, K. Ecker, N. Garritsen, H. Adv Hematol Research Article Background. We developed a completely closed system based on gravity separation without centrifugation steps for separation of whole blood. With this new system we compared quality and stability of the processed blood components (PRC and plasma) with respect to classical preparation. Furthermore the cost-effectiveness of this hollow fibre system was evaluated. Study Design and Methods. Whole blood collections of 15 regular blood donors were used for component preparation using the U shaped hollow fibre filter device. Results were compared to 15 whole blood preparations using centrifugation. The following parameters were evaluated: total hemoglobin, leukocyte counts, the serum concentration of total protein, lactate dehydrogenase (LDH) and potassium. Furthermore ATIII, vWF and F VIII were analyzed at different timepoints. Results. packed red cells: the data directly after separation and after 42 days of storage are in line with the guidelines of the council of Europe. Plasma. all plasma quality data are in line with the guidelines of the council of Europe for quality assurance of plasma, except for a low protein amount (factor 0.75). Conclusion. Separation of whole blood on a clinical scale in this new closed system is feasible, however the plasma protein content must be optimized. Hindawi Publishing Corporation 2009 2010-02-04 /pmc/articles/PMC2821640/ /pubmed/20169121 http://dx.doi.org/10.1155/2009/175234 Text en Copyright © 2009 T. Brune et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Brune, T. Hannemann-Pohl, K. Nißle, K. Ecker, N. Garritsen, H. Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device |
title | Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device |
title_full | Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device |
title_fullStr | Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device |
title_full_unstemmed | Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device |
title_short | Quality, Stability, and Safety Data of Packed Red Cells and Plasma Processed by Gravity Separation Using a New Fully Integrated Hollow-Fibre Filter Device |
title_sort | quality, stability, and safety data of packed red cells and plasma processed by gravity separation using a new fully integrated hollow-fibre filter device |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2821640/ https://www.ncbi.nlm.nih.gov/pubmed/20169121 http://dx.doi.org/10.1155/2009/175234 |
work_keys_str_mv | AT brunet qualitystabilityandsafetydataofpackedredcellsandplasmaprocessedbygravityseparationusinganewfullyintegratedhollowfibrefilterdevice AT hannemannpohlk qualitystabilityandsafetydataofpackedredcellsandplasmaprocessedbygravityseparationusinganewfullyintegratedhollowfibrefilterdevice AT nißlek qualitystabilityandsafetydataofpackedredcellsandplasmaprocessedbygravityseparationusinganewfullyintegratedhollowfibrefilterdevice AT eckern qualitystabilityandsafetydataofpackedredcellsandplasmaprocessedbygravityseparationusinganewfullyintegratedhollowfibrefilterdevice AT garritsenh qualitystabilityandsafetydataofpackedredcellsandplasmaprocessedbygravityseparationusinganewfullyintegratedhollowfibrefilterdevice |