Cargando…

Review of Adverse Events Associated With False Glucose Readings Measured by GDH-PQQ–Based Glucose Test Strips in the Presence of Interfering Sugars

OBJECTIVE: To assess the implications of falsely elevated glucose readings measured with glucose dehydrogenase pyrroloquinolinequinone (GDH-PQQ) test strips. RESEARCH DESIGN AND METHODS: We conducted a review of the Food and Drug Administration's Manufacturer and User Facility Device Experience...

Descripción completa

Detalles Bibliográficos
Autores principales: Frias, Juan P., Lim, Christine G., Ellison, John M., Montandon, Carol M.
Formato: Texto
Lenguaje:English
Publicado: American Diabetes Association 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2845013/
https://www.ncbi.nlm.nih.gov/pubmed/20351227
http://dx.doi.org/10.2337/dc09-1822
Descripción
Sumario:OBJECTIVE: To assess the implications of falsely elevated glucose readings measured with glucose dehydrogenase pyrroloquinolinequinone (GDH-PQQ) test strips. RESEARCH DESIGN AND METHODS: We conducted a review of the Food and Drug Administration's Manufacturer and User Facility Device Experience database and medical literature for adverse events (AEs) associated with falsely elevated glucose readings with GDH-PQQ test strips in the presence of interfering sugars. RESULTS: Eighty-two reports were identified: 16 (20%) were associated with death, 46 (56%) with severe hypoglycemia, and 12 (15%) with nonsevere hypoglycemia. In eight reports (10%), the AE was not described. Forty-two events (51%) occurred in the U.S. Although most events occurred in hospitalized patients, at least 14 (17%) occurred in outpatients. Agents most commonly associated with AEs were icodextrin-containing peritoneal dialysate and maltose-containing intravenous immune globulin. CONCLUSIONS: GDH-PQQ test strips pose a safety risk to insulin-using patients treated with agents containing or metabolized to interfering sugars.