Cargando…

Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control

The automated slidemaker/stainers of the four Beckman Coulter LH755 hematology systems in our laboratory are operated as analyzers, with similar requirements for setup, maintenance and quality control. A study was performed to confirm that these slide maker/stainers in routine use produce peripheral...

Descripción completa

Detalles Bibliográficos
Autores principales: SIMSON, E, GASCON-LEMA, M G, BROWN, D L
Formato: Texto
Lenguaje:English
Publicado: Blackwell Publishing Ltd 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2847201/
https://www.ncbi.nlm.nih.gov/pubmed/19220552
http://dx.doi.org/10.1111/j.1751-553X.2009.01141.x
_version_ 1782179547001651200
author SIMSON, E
GASCON-LEMA, M G
BROWN, D L
author_facet SIMSON, E
GASCON-LEMA, M G
BROWN, D L
author_sort SIMSON, E
collection PubMed
description The automated slidemaker/stainers of the four Beckman Coulter LH755 hematology systems in our laboratory are operated as analyzers, with similar requirements for setup, maintenance and quality control. A study was performed to confirm that these slide maker/stainers in routine use produce peripheral blood films that are completely satisfactory for microscopy and without cells, particularly abnormal cells, being pulled to the edges or sides of the film outside the usual working area. One hundred and thirty-nine automated blood films that had been produced during routine operation were compared with well-prepared manual films from the same patients. None of the films was unacceptable for microscopy. The distributions of normal white cell types within the counting areas of automated films compared with manual films, for all 139 samples for WBC from 1.0 to 352.8 × 10(9)/l; for blasts and promyelocytes in the 65 samples in which they occurred and for nucleated red blood cells in the 58 samples in which they occurred all fell within the expected limits of 200 cell differential counts of CLSI H20-A. Red cell morphology and the occurrence of WBC clumps, platelet clumps and smudge cells were comparable between the automated and manual films of all samples. We conclude that automated slidemaker/stainers, as typified by those of the Beckman Coulter LH755 system, are capable of producing blood films comparable with well-prepared manual films in routine laboratory use; and that the maintenance and quality control procedures used in our laboratory ensure consistent high quality performance from these systems.
format Text
id pubmed-2847201
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Blackwell Publishing Ltd
record_format MEDLINE/PubMed
spelling pubmed-28472012010-04-08 Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control SIMSON, E GASCON-LEMA, M G BROWN, D L Int J Lab Hematol Original Articles The automated slidemaker/stainers of the four Beckman Coulter LH755 hematology systems in our laboratory are operated as analyzers, with similar requirements for setup, maintenance and quality control. A study was performed to confirm that these slide maker/stainers in routine use produce peripheral blood films that are completely satisfactory for microscopy and without cells, particularly abnormal cells, being pulled to the edges or sides of the film outside the usual working area. One hundred and thirty-nine automated blood films that had been produced during routine operation were compared with well-prepared manual films from the same patients. None of the films was unacceptable for microscopy. The distributions of normal white cell types within the counting areas of automated films compared with manual films, for all 139 samples for WBC from 1.0 to 352.8 × 10(9)/l; for blasts and promyelocytes in the 65 samples in which they occurred and for nucleated red blood cells in the 58 samples in which they occurred all fell within the expected limits of 200 cell differential counts of CLSI H20-A. Red cell morphology and the occurrence of WBC clumps, platelet clumps and smudge cells were comparable between the automated and manual films of all samples. We conclude that automated slidemaker/stainers, as typified by those of the Beckman Coulter LH755 system, are capable of producing blood films comparable with well-prepared manual films in routine laboratory use; and that the maintenance and quality control procedures used in our laboratory ensure consistent high quality performance from these systems. Blackwell Publishing Ltd 2010-02 /pmc/articles/PMC2847201/ /pubmed/19220552 http://dx.doi.org/10.1111/j.1751-553X.2009.01141.x Text en Journal compilation © 2010 Blackwell Publishing Ltd http://creativecommons.org/licenses/by/2.5/ Re-use of this article is permitted in accordance with the Creative Commons Deed, Attribution 2.5, which does not permit commercial exploitation.
spellingShingle Original Articles
SIMSON, E
GASCON-LEMA, M G
BROWN, D L
Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control
title Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control
title_full Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control
title_fullStr Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control
title_full_unstemmed Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control
title_short Performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control
title_sort performance of automated slidemakers and stainers in a working laboratory environment – routine operation and quality control
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2847201/
https://www.ncbi.nlm.nih.gov/pubmed/19220552
http://dx.doi.org/10.1111/j.1751-553X.2009.01141.x
work_keys_str_mv AT simsone performanceofautomatedslidemakersandstainersinaworkinglaboratoryenvironmentroutineoperationandqualitycontrol
AT gasconlemamg performanceofautomatedslidemakersandstainersinaworkinglaboratoryenvironmentroutineoperationandqualitycontrol
AT browndl performanceofautomatedslidemakersandstainersinaworkinglaboratoryenvironmentroutineoperationandqualitycontrol