Comparability of Results between Point-of-Care and Automated Instruments to Measure B-type Natriuretic Peptide

OBJECTIVES: Heart failure is one of the leading causes of death in the U.S. The incorporation of B-type natriuretic peptide (BNP) measurements when triaging patients presenting with shortness of breath has improved the diagnostic and prognostic ability of physicians. Currently, there are no point-of...

Descripción completa

Detalles Bibliográficos
Autores principales: Shah, Kevin, Terracciano, Garrett J., Jiang, Kevin, Maisel, Alan S., Fitzgerald, Robert L.
Formato: Texto
Lenguaje:English
Publicado: Department of Emergency Medicine, University of California, Irvine School of Medicine 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2850853/
https://www.ncbi.nlm.nih.gov/pubmed/20411075
Descripción
Sumario:OBJECTIVES: Heart failure is one of the leading causes of death in the U.S. The incorporation of B-type natriuretic peptide (BNP) measurements when triaging patients presenting with shortness of breath has improved the diagnostic and prognostic ability of physicians. Currently, there are no point-of-care systems for quantifying BNP that can be used without sacrificing accuracy. We compared the analytical performance of the Abbott i-STAT analyzer, a handheld point-of-care system for measuring BNP, with the lab-based system, the Abbott ARCHITECT. METHODS: One-hundred fifty samples were collected from three clinical settings: 41 from the Emergency Department, 58 from the inpatient wards, and 51 from heart failure outpatient clinics. Linear regression and bias difference analyses were run to evaluate the accuracy of the i-STAT. Correlation between the i-STAT and Architect BNP values were made with values of BNP. RESULTS: The correlation coefficient was r=0.977 (N=150, p<.0001). The average bias was significant (-36) and there were concentration-dependent differences at higher BNP values. Precision of the i-STAT was poor compared to the lab-based platform. CONCLUSION: Although the precision of the i-STAT was poor, there was good clinical agreement between the i-STAT and the lab-based platform.