Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans
BACKGROUND: We developed a cardiac SPECT system (X-ACT) with low dose volume CT transmission-based attenuation correction (AC). Three solid-state detectors are configured to form a triple-head system for emission scans and reconfigured to form a 69-cm field-of-view detector arc for transmission scan...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Springer-Verlag
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2866957/ https://www.ncbi.nlm.nih.gov/pubmed/20169476 http://dx.doi.org/10.1007/s12350-010-9204-8 |
_version_ | 1782180950304620544 |
---|---|
author | Bai, Chuanyong Conwell, Richard Kindem, Joel Babla, Hetal Gurley, Mike De Los Santos, Romer Old, Rex Weatherhead, Randy Arram, Samia Maddahi, Jamshid |
author_facet | Bai, Chuanyong Conwell, Richard Kindem, Joel Babla, Hetal Gurley, Mike De Los Santos, Romer Old, Rex Weatherhead, Randy Arram, Samia Maddahi, Jamshid |
author_sort | Bai, Chuanyong |
collection | PubMed |
description | BACKGROUND: We developed a cardiac SPECT system (X-ACT) with low dose volume CT transmission-based attenuation correction (AC). Three solid-state detectors are configured to form a triple-head system for emission scans and reconfigured to form a 69-cm field-of-view detector arc for transmission scans. A near mono-energetic transmission line source is produced from the collimated fluorescence x-ray emitted from a lead target when the target is illuminated by a narrow polychromatic x-ray beam from an x-ray tube. Transmission scans can be completed in 1 min with insignificant patient dose (deep dose equivalent <5 μSv). METHODS: We used phantom studies to evaluate (1) the accuracy of the reconstructed attenuation maps, (2) the effect of AC on image uniformity, and (3) the effect of AC on defect contrast (DC). The phantoms we used included an ACR phantom, an anthropomorphic phantom with a uniform cardiac insert, and an anthropomorphic phantom with two defects in the cardiac insert. RESULTS: The reconstructed attenuation coefficient of water at 140 keV was .150 ± .003/cm in the uniform region of the ACR phantom, .151 ± .003/cm and .151 ± .002/cm in the liver and cardiac regions of the anthropomorphic phantom. The ACR phantom images with AC showed correction of the bowing effect due to attenuation in the images without AC (NC). The 17-segment scores of the images of the uniform cardiac insert were 78.3 ± 6.5 before and 87.9 ± 3.3 after AC (average ± standard deviation). The inferior-to-anterior wall ratio and the septal-to-lateral wall ratio were .99 and 1.16 before and 1.02 and 1.00 after AC. The DC of the two defects was .528 and .156 before and .628 and .173 after AC. CONCLUSION: The X-ACT system generated accurate attenuation maps with 1-minute transmission scans. AC improved image quality and uniformity over NC. |
format | Text |
id | pubmed-2866957 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Springer-Verlag |
record_format | MEDLINE/PubMed |
spelling | pubmed-28669572010-05-24 Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans Bai, Chuanyong Conwell, Richard Kindem, Joel Babla, Hetal Gurley, Mike De Los Santos, Romer Old, Rex Weatherhead, Randy Arram, Samia Maddahi, Jamshid J Nucl Cardiol Original Article BACKGROUND: We developed a cardiac SPECT system (X-ACT) with low dose volume CT transmission-based attenuation correction (AC). Three solid-state detectors are configured to form a triple-head system for emission scans and reconfigured to form a 69-cm field-of-view detector arc for transmission scans. A near mono-energetic transmission line source is produced from the collimated fluorescence x-ray emitted from a lead target when the target is illuminated by a narrow polychromatic x-ray beam from an x-ray tube. Transmission scans can be completed in 1 min with insignificant patient dose (deep dose equivalent <5 μSv). METHODS: We used phantom studies to evaluate (1) the accuracy of the reconstructed attenuation maps, (2) the effect of AC on image uniformity, and (3) the effect of AC on defect contrast (DC). The phantoms we used included an ACR phantom, an anthropomorphic phantom with a uniform cardiac insert, and an anthropomorphic phantom with two defects in the cardiac insert. RESULTS: The reconstructed attenuation coefficient of water at 140 keV was .150 ± .003/cm in the uniform region of the ACR phantom, .151 ± .003/cm and .151 ± .002/cm in the liver and cardiac regions of the anthropomorphic phantom. The ACR phantom images with AC showed correction of the bowing effect due to attenuation in the images without AC (NC). The 17-segment scores of the images of the uniform cardiac insert were 78.3 ± 6.5 before and 87.9 ± 3.3 after AC (average ± standard deviation). The inferior-to-anterior wall ratio and the septal-to-lateral wall ratio were .99 and 1.16 before and 1.02 and 1.00 after AC. The DC of the two defects was .528 and .156 before and .628 and .173 after AC. CONCLUSION: The X-ACT system generated accurate attenuation maps with 1-minute transmission scans. AC improved image quality and uniformity over NC. Springer-Verlag 2010-02-19 2010 /pmc/articles/PMC2866957/ /pubmed/20169476 http://dx.doi.org/10.1007/s12350-010-9204-8 Text en © The Author(s) 2010 https://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited. |
spellingShingle | Original Article Bai, Chuanyong Conwell, Richard Kindem, Joel Babla, Hetal Gurley, Mike De Los Santos, Romer Old, Rex Weatherhead, Randy Arram, Samia Maddahi, Jamshid Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans |
title | Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans |
title_full | Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans |
title_fullStr | Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans |
title_full_unstemmed | Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans |
title_short | Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans |
title_sort | phantom evaluation of a cardiac spect/vct system that uses a common set of solid-state detectors for both emission and transmission scans |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2866957/ https://www.ncbi.nlm.nih.gov/pubmed/20169476 http://dx.doi.org/10.1007/s12350-010-9204-8 |
work_keys_str_mv | AT baichuanyong phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT conwellrichard phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT kindemjoel phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT bablahetal phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT gurleymike phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT delossantosromer phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT oldrex phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT weatherheadrandy phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT arramsamia phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans AT maddahijamshid phantomevaluationofacardiacspectvctsystemthatusesacommonsetofsolidstatedetectorsforbothemissionandtransmissionscans |