Cargando…

Separation of the Tip of a Coblation Wand within the Knee Joint: A Complication of Arthroscopic Adhesiolysis

Coblation devices are now widely used in arthroscopic surgery and they show a very low incidence of intraoperative complications. We experienced a case where the tip of the wand separated and migrated into the posterior knee compartment in an arthrofibrotic knee. The free wand tip was identified and...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Yeub, Kim, Ha-Kyung, Yoon, Jung-Ro, Muzaffar, Nasir, Kim, Taik-Sun, Shin, Young-Su
Formato: Texto
Lenguaje:English
Publicado: The Korean Orthopaedic Association 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2867198/
https://www.ncbi.nlm.nih.gov/pubmed/20514271
http://dx.doi.org/10.4055/cios.2010.2.2.125
Descripción
Sumario:Coblation devices are now widely used in arthroscopic surgery and they show a very low incidence of intraoperative complications. We experienced a case where the tip of the wand separated and migrated into the posterior knee compartment in an arthrofibrotic knee. The free wand tip was identified and then extricated from the popliteal hiatus of the knee with using C-arm fluoroscopic control. To the best of our knowledge, this is the first report of its kind involving coblation wands. We describe this complication to show that the use of coblation devices can lead to unexpected problems and it is imperative to inspect all instruments before and after each surgical use.