Real-Time Hypoglycemia Prediction Suite Using Continuous Glucose Monitoring: A safety net for the artificial pancreas

OBJECTIVE: The purpose of this study was to develop an advanced algorithm that detects pending hypoglycemia and then suspends basal insulin delivery. This approach can provide a solution to the problem of nocturnal hypoglycemia, a major concern of patients with diabetes. RESEARCH DESIGN AND METHODS:...

Descripción completa

Detalles Bibliográficos
Autores principales: Dassau, Eyal, Cameron, Fraser, Lee, Hyunjin, Bequette, B. Wayne, Zisser, Howard, Jovanovič, Lois, Chase, H. Peter, Wilson, Darrell M., Buckingham, Bruce A., Doyle, Francis J.
Formato: Texto
Lenguaje:English
Publicado: American Diabetes Association 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2875433/
https://www.ncbi.nlm.nih.gov/pubmed/20508231
http://dx.doi.org/10.2337/dc09-1487
Descripción
Sumario:OBJECTIVE: The purpose of this study was to develop an advanced algorithm that detects pending hypoglycemia and then suspends basal insulin delivery. This approach can provide a solution to the problem of nocturnal hypoglycemia, a major concern of patients with diabetes. RESEARCH DESIGN AND METHODS: This real-time hypoglycemia prediction algorithm (HPA) combines five individual algorithms, all based on continuous glucose monitoring 1-min data. A predictive alarm is issued by a voting algorithm when a hypoglycemic event is predicted to occur in the next 35 min. The HPA system was developed using data derived from 21 Navigator studies that assessed Navigator function over 24 h in children with type 1 diabetes. We confirmed the function of the HPA using a separate dataset from 22 admissions of type 1 diabetic subjects. During these admissions, hypoglycemia was induced by gradual increases in the basal insulin infusion rate up to 180% from the subject's own baseline infusion rate. RESULTS: Using a prediction horizon of 35 min, a glucose threshold of 80 mg/dl, and a voting threshold of three of five algorithms to predict hypoglycemia (defined as a FreeStyle plasma glucose readings <60 mg/dl), the HPA predicted 91% of the hypoglycemic events. When four of five algorithms were required to be positive, then 82% of the events were predicted. CONCLUSIONS: The HPA will enable automated insulin-pump suspension in response to a pending event that has been detected prior to severe immediate complications.