Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference (137)Cs brachytherapy source and re...
Autores principales: | , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2884301/ https://www.ncbi.nlm.nih.gov/pubmed/20589119 http://dx.doi.org/10.4103/0971-6203.62200 |
_version_ | 1782182317349928960 |
---|---|
author | Vandana, S. Sharma, S. D. |
author_facet | Vandana, S. Sharma, S. D. |
author_sort | Vandana, S. |
collection | PubMed |
description | Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference (137)Cs brachytherapy source and recommend an optimum frequency of recalibration. An HDR-1000 Plus well-type ionization chamber, a reference (137)Cs brachytherapy source (CDCSJ5), and a MAX-4000 electrometer were used in this study. The HDR-1000 Plus well-type chamber was calibrated in terms of reference air kerma rate by the Standards Laboratory of the International Atomic Energy Agency (IAEA), Vienna. The response of the chamber was verified at regular intervals over a period of eight years using the reference (137)Cs source. All required correction factors were applied in the calculation of the RAKR of the (137)Cs source. This study reveals that the response of the HDR-1000 Plus well-type chamber was well within ±0.5% for about three years after calibration/recalibration. However, it shows deviations larger than ±0.5% after three years of calibration/recalibration and the maximum variation in response of the chamber during an eight year period was 1.71%. The optimum frequency of recalibration of a high dose rate well-type chamber should be three years. |
format | Text |
id | pubmed-2884301 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Medknow Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-28843012010-06-29 Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources Vandana, S. Sharma, S. D. J Med Phys Original Article Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference (137)Cs brachytherapy source and recommend an optimum frequency of recalibration. An HDR-1000 Plus well-type ionization chamber, a reference (137)Cs brachytherapy source (CDCSJ5), and a MAX-4000 electrometer were used in this study. The HDR-1000 Plus well-type chamber was calibrated in terms of reference air kerma rate by the Standards Laboratory of the International Atomic Energy Agency (IAEA), Vienna. The response of the chamber was verified at regular intervals over a period of eight years using the reference (137)Cs source. All required correction factors were applied in the calculation of the RAKR of the (137)Cs source. This study reveals that the response of the HDR-1000 Plus well-type chamber was well within ±0.5% for about three years after calibration/recalibration. However, it shows deviations larger than ±0.5% after three years of calibration/recalibration and the maximum variation in response of the chamber during an eight year period was 1.71%. The optimum frequency of recalibration of a high dose rate well-type chamber should be three years. Medknow Publications 2010 /pmc/articles/PMC2884301/ /pubmed/20589119 http://dx.doi.org/10.4103/0971-6203.62200 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Vandana, S. Sharma, S. D. Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources |
title | Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources |
title_full | Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources |
title_fullStr | Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources |
title_full_unstemmed | Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources |
title_short | Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources |
title_sort | long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2884301/ https://www.ncbi.nlm.nih.gov/pubmed/20589119 http://dx.doi.org/10.4103/0971-6203.62200 |
work_keys_str_mv | AT vandanas longtermresponsestabilityofawelltypeionizationchamberusedincalibrationofhighdoseratebrachytherapysources AT sharmasd longtermresponsestabilityofawelltypeionizationchamberusedincalibrationofhighdoseratebrachytherapysources |