Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources

Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference (137)Cs brachytherapy source and re...

Descripción completa

Detalles Bibliográficos
Autores principales: Vandana, S., Sharma, S. D.
Formato: Texto
Lenguaje:English
Publicado: Medknow Publications 2010
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2884301/
https://www.ncbi.nlm.nih.gov/pubmed/20589119
http://dx.doi.org/10.4103/0971-6203.62200
_version_ 1782182317349928960
author Vandana, S.
Sharma, S. D.
author_facet Vandana, S.
Sharma, S. D.
author_sort Vandana, S.
collection PubMed
description Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference (137)Cs brachytherapy source and recommend an optimum frequency of recalibration. An HDR-1000 Plus well-type ionization chamber, a reference (137)Cs brachytherapy source (CDCSJ5), and a MAX-4000 electrometer were used in this study. The HDR-1000 Plus well-type chamber was calibrated in terms of reference air kerma rate by the Standards Laboratory of the International Atomic Energy Agency (IAEA), Vienna. The response of the chamber was verified at regular intervals over a period of eight years using the reference (137)Cs source. All required correction factors were applied in the calculation of the RAKR of the (137)Cs source. This study reveals that the response of the HDR-1000 Plus well-type chamber was well within ±0.5% for about three years after calibration/recalibration. However, it shows deviations larger than ±0.5% after three years of calibration/recalibration and the maximum variation in response of the chamber during an eight year period was 1.71%. The optimum frequency of recalibration of a high dose rate well-type chamber should be three years.
format Text
id pubmed-2884301
institution National Center for Biotechnology Information
language English
publishDate 2010
publisher Medknow Publications
record_format MEDLINE/PubMed
spelling pubmed-28843012010-06-29 Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources Vandana, S. Sharma, S. D. J Med Phys Original Article Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference (137)Cs brachytherapy source and recommend an optimum frequency of recalibration. An HDR-1000 Plus well-type ionization chamber, a reference (137)Cs brachytherapy source (CDCSJ5), and a MAX-4000 electrometer were used in this study. The HDR-1000 Plus well-type chamber was calibrated in terms of reference air kerma rate by the Standards Laboratory of the International Atomic Energy Agency (IAEA), Vienna. The response of the chamber was verified at regular intervals over a period of eight years using the reference (137)Cs source. All required correction factors were applied in the calculation of the RAKR of the (137)Cs source. This study reveals that the response of the HDR-1000 Plus well-type chamber was well within ±0.5% for about three years after calibration/recalibration. However, it shows deviations larger than ±0.5% after three years of calibration/recalibration and the maximum variation in response of the chamber during an eight year period was 1.71%. The optimum frequency of recalibration of a high dose rate well-type chamber should be three years. Medknow Publications 2010 /pmc/articles/PMC2884301/ /pubmed/20589119 http://dx.doi.org/10.4103/0971-6203.62200 Text en © Journal of Medical Physics http://creativecommons.org/licenses/by/2.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Vandana, S.
Sharma, S. D.
Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
title Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
title_full Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
title_fullStr Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
title_full_unstemmed Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
title_short Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
title_sort long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2884301/
https://www.ncbi.nlm.nih.gov/pubmed/20589119
http://dx.doi.org/10.4103/0971-6203.62200
work_keys_str_mv AT vandanas longtermresponsestabilityofawelltypeionizationchamberusedincalibrationofhighdoseratebrachytherapysources
AT sharmasd longtermresponsestabilityofawelltypeionizationchamberusedincalibrationofhighdoseratebrachytherapysources