Optimum b value for resolving crossing fibers: a study with standard clinical b value using 1.5-T MR
INTRODUCTION: We sought to investigate the optimum b value for resolving crossing fiber using high-angular resolution diffusion imaging (HARDI)-based multi-tensor tractography. The study tested the standard b values that are commonly used in the routine clinical setting. METHODS: Ten normal voluntee...
Autores principales: | , , , , , |
---|---|
Formato: | Texto |
Lenguaje: | English |
Publicado: |
Springer-Verlag
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2901494/ https://www.ncbi.nlm.nih.gov/pubmed/20309533 http://dx.doi.org/10.1007/s00234-010-0670-0 |
Sumario: | INTRODUCTION: We sought to investigate the optimum b value for resolving crossing fiber using high-angular resolution diffusion imaging (HARDI)-based multi-tensor tractography. The study tested the standard b values that are commonly used in the routine clinical setting. METHODS: Ten normal volunteers (five men and five women) with a mean age of 26.3 years (range, 22–32 years) were scanned using a 1.5-T clinical magnetic resonance unit. Single-shot echo-planar imaging was used for diffusion-weighted imaging with a diffusion-sensitizing gradient in 32 orientations. The b values of 700, 1,400, 2,100, and 2,800 s/m(2) were used. Data postprocessing was performed using multi-tensor methods. The depiction of the optic nerves, optic tracts, and decussation of superior cerebellar peduncles were assessed. RESULTS: The depictions of the nerve fibers were independent of the b values tested. CONCLUSION: The depiction of crossing fibers by HARDI-based multi-tensor tractography is not substantially influenced by b values ranging from 700 to 2,800 s/m(2). Thus, the optimum b value within this range may be the lowest one considering the higher signal to noise ratio. |
---|